Measurement Performance
Example | The flowmeter is installed and proved for the first time. The flowmeter |
| mass measurement is 250.27 lb; the reference device measurement is |
| 250 lb. A mass flow meter factor is determined as follows: |
| 250 |
| MassFlowMeterFactor = 1 ⋅ |
| 250.27 |
| The first mass flow meter factor is 0.9989. |
| One year later, the flowmeter is proved again. The flowmeter mass |
| measurement is 250.07 lb; the reference device measurement is |
| 250.25 lb. A new mass flow meter factor is determined as follows: |
| MassFlowMeterFactor = 0.9989 ⋅ |
| 250.07 |
| The new mass flow meter factor is 0.9996. |
|
|
10.5Performing density calibration
Density calibration includes the following calibration points:
•All sensors:
-D1 calibration
-D2 calibration
•
-D3 calibration (optional)
-D4 calibration (optional)
For
•Do not perform the D1 or D2 calibration.
•Perform D3 calibration if you have one calibrated fluid.
•Perform both D3 and D4 calibrations if you have two calibrated fluids (other than air and water).
The calibrations that you choose must be performed without interruption, in the order listed here.
Note: Before performing the calibration, record your current calibration parameters. If you are using ProLink II, you can do this by saving the current configuration to a file on the PC. If the calibration fails, restore the known values.
You can calibrate for density with ProLink II.
10.5.1Preparing for density calibration
Before beginning density calibration, review the requirements in this section.
Sensor requirements
During density calibration, the sensor must be completely filled with the calibration fluid, and flow through the sensor must be at the lowest rate allowed by your application. This is usually accomplished by closing the shutoff valve downstream from the sensor, then filling the sensor with the appropriate fluid.
Compensation
Measurement Performance
Troubleshooting
Defaults
Configuration and Use Manual | 87 |