Measurement Performance
10.2.4Comparison and recommendations
When choosing among meter verification, meter validation, and calibration, consider the following factors:
•Process interruption
-Meter verification requires approximately four minutes to perform. During these four minutes, flow can continue (provided sufficient stability is maintained); however, outputs will not report process data.
-Meter validation for density does not interrupt the process at all. However, meter validation for mass flow or volume flow requires process
-Calibration requires process
•External measurement requirements
-Meter verification does not require external measurements.
-Zero calibration does not require external measurements.
-Density calibration, temperature calibration, and meter validation require external measurements. For good results, the external measurement must be highly accurate.
•Measurement adjustment
-Meter verification is an indicator of sensor condition, but does not change flowmeter internal measurement in any way.
-Meter validation does not change flowmeter internal measurement in any way. If you decide to adjust a meter factor as a result of a meter validation procedure, only the reported measurement is changed – the base measurement is not changed. You can always reverse the change by returning the meter factor to its previous value.
-Calibration changes the transmitter’s interpretation of process data, and accordingly changes the base measurement. If you perform a zero calibration, you can restore the factory zero at a later time. You cannot return to the previous zero (if different from the factory zero), density calibration values, or temperature calibration values unless you have manually recorded them.
Micro Motion recommends obtaining the meter verification transmitter option and performing meter verification on a regular basis.
10.3Performing meter verification
Note: To use meter verification, the transmitter must be paired with an enhanced core processor, and the meter verification option must be purchased for the transmitter.
The meter verification procedure can be performed on any process fluid. It is not necessary to match factory conditions. Meter verification is not affected by any parameters configured for flow, density, or temperature.
During the test, process conditions must be stable. To maximize stability:
•Maintain a constant temperature and pressure.
•Avoid changes to fluid composition (e.g.,
•Maintain a constant flow. For higher test certainty, reduce or stop flow.
Compensation
Measurement Performance
Troubleshooting
Defaults
Configuration and Use Manual | 83 |