Configuring the Filling and Dosing Application
Another AOC calibration is recommended:
•If equipment has been replaced or adjusted
•If flow rate has changed significantly
•If fills are consistently missing the target value
7.5.3Rolling AOC calibration
Note: In common use, the first fill may be slightly overfilled because the default compensation factor is 0.2. To prevent this, increase the AOC Coeff value in the Run Filler window (see Figure
To enable rolling AOC calibration:
1.Click ProLink > Run Filler. The window shown in Figure
2.Click Start AOC Cal. The AOC Calibration Active light turns red.
3.Begin filling. Do not click Save AOC Cal. The AOC coefficient is recalculated after each fill, and the current value is displayed in the Run Filler window.
At any time, you can click Save AOC Cal. The current AOC coefficient will be saved in the transmitter and used for all overshoot compensation during subsequent fills. In other words, this action changes the AOC calibration method from rolling to standard.
Using the Transmitter
Optional Configuration
Filler Configuration
Using the Filler
Configuration and Use Manual | 65 |