10.6Performing temperature calibration

Measurement Performance

10.6Performing temperature calibration

Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope calibration. The entire procedure must be completed without interruption.

You can calibrate for temperature with ProLink II. See Figure 10-4.

Figure 10-4Temperature calibration – ProLink II

Temperature Offset calibration

Fill sensor with low-

temperature fluid

Wait until sensor achieves

thermal equilibrium

ProLink Menu >

Calibration >

Temp offset cal

Enter temperature of low-

temperature fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

Temperature Slope calibration

Fill sensor with high-

temperature fluid

Wait until sensor achieves

thermal equilibrium

ProLink Menu >

Calibration >

Temp slope cal

Enter temperature of high-

temperature fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

Done

90

Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application

Page 98