![Density calibration fluids](/images/new-backgrounds/15629/15629191x1.webp)
Measurement Performance
Density calibration fluids
D1 and D2 density calibration require a D1
CAUTION
For
For D3 density calibration, the D3 fluid must meet the following requirements:
•Minimum density of 0.6 g/cm3
•Minimum difference of 0.1 g/cm3 between the density of the D3 fluid and the density of water. The density of the D3 fluid may be either greater or less than the density of water
For D4 density calibration, the D4 fluid must meet the following requirements:
•Minimum density of 0.6 g/cm3
•Minimum difference of 0.1 g/cm3 between the density of the D4 fluid and the density of the D3 fluid. The density of the D4 fluid must be greater than the density of the D3 fluid
•Minimum difference of 0.1 g/cm3 between the density of the D4 fluid and the density of water. The density of the D4 fluid may be either greater or less than the density of water
10.5.2Density calibration procedures
To perform a D1 and D2 density calibration, see Figure
To perform a D3 density calibration or a D3 and D4 density calibration, see Figure
88 | Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application |