Emerson Process Management 1500 Density calibration fluids, 10.5.2Density calibration procedures

Models: 1500

1 146
Download 146 pages 48.94 Kb
Page 96
Image 96
Density calibration fluids

Measurement Performance

Density calibration fluids

D1 and D2 density calibration require a D1 (low-density) fluid and a D2 (high-density) fluid. You may use air and water. If you are calibrating a T-Series sensor, the D1 fluid must be air and the D2 fluid must be water.

10.5.2Density calibration proceduresManual backgroundManual background CAUTION

For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be performed on water.

For D3 density calibration, the D3 fluid must meet the following requirements:

Minimum density of 0.6 g/cm3

Minimum difference of 0.1 g/cm3 between the density of the D3 fluid and the density of water. The density of the D3 fluid may be either greater or less than the density of water

For D4 density calibration, the D4 fluid must meet the following requirements:

Minimum density of 0.6 g/cm3

Minimum difference of 0.1 g/cm3 between the density of the D4 fluid and the density of the D3 fluid. The density of the D4 fluid must be greater than the density of the D3 fluid

Minimum difference of 0.1 g/cm3 between the density of the D4 fluid and the density of water. The density of the D4 fluid may be either greater or less than the density of water

10.5.2Density calibration procedures

To perform a D1 and D2 density calibration, see Figure 10-2.

To perform a D3 density calibration or a D3 and D4 density calibration, see Figure 10-3.

88

Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application

Page 96
Image 96
Emerson Process Management 1500 manual Density calibration fluids, 10.5.2Density calibration procedures