CONTROL FUNCTIONS 4-15
MON2000
JULY 2010 Stop Now
provided scroll arrows to increase or
decrease the highlighted value.
5. Click the button or press the
enter key to write the selected date/time to
the GC Controller.
Click the button to abort and
return to the main window.
4.8 STOP NOW
Use this function to immediately stop all
analysis runs.
1. Use the Control > Stop Now menu to access
this function.
2. MON2000 prompts you to confirm your
selection.
Click the button to proceed.
Click the button to abort and return
to the main window.
This function forces the system to the Idle mode. If
Stop Now is performed while an analysis is in
progress, the components will continue to elute from
the columns per the Idle mode. All resultant data will
be lost.
Do not perform a Stop Now unless absolutely
necessary. Whenever possible, use the Halt function.