APPLICATION FUNCTIONS 5-15
MON2000
JULY 2010 Timed Events
Valve Events
Type - use the pull-down menu to select
between valve number (Valve #), discrete
output channel number (DO #), and stream
switch (Strm Sw); additional selections
SSO1, SSO2, Bleed1, Bleed2, Block1 and
Block2 are available for the Model 700 GC
applications
Valve/DO # (for valves and discrete outputs
only) - input the valve number or discrete
output channel number. Use the provided
combo box to select valve number by name
State (for all types except for stream switch)
- use the pull-down menu to toggle ON\OFF
Time - input the time that the event is to
occur. LIMITS: 0.0 to 6000.0 seconds
Integration Events
Type - use the pull-down menu to select
between Inhibit, Integrate, Auto Zero, Slope
Sensitivity, Peak Width, Single Baseline,
FID Gain, Summation, or Fused Override
Value - For peak width or slope sensitivity,
input the number of points to be used.
LIMITS: Peak Width and Slope Sensitivity,
1 to 99
For FID gain, use the pull-down menu to
toggle Low\High.
For Single Baseline, use the pull-down
menu to toggle OFF\BGN (Begin-
ning)\END.
Except for AutoZero, for all others use the
pull-down menu to toggle ON\OFF. Auto
Zero does not have 'Value' field.
Det - enter the appropriate detector number
that the event will effect. Valid range is '1'
or '2'
Time - input the time that the event is to
occur. LIMITS: 0.0 to 6000.0 seconds