5-40 APPLICATION FUNCTIONS
MON2000
Streams JULY 2010
The standard settings include:
Table 5-4 Description of Standard Stream Settings
Std Setting Description
<Number> number label assigned by MON2000
This number correlates with the stream
switch. A maximum of 12 streams can be con-
trolled by MON2000 and a standard GC appli-
cation.
Name text string label to identify a stream (12
character maximum)
Use how stream is used
Settings include: calibration (CAL); analysis
(ANL); not used (NOT)
Det detector used by this stream
Control of the detector is application-specific.
For example,
• for Model 700/2350A, Modes 0 and 1, the
Det col is “DET1” and read-only
• for Model 700/2350A, Mode 2, the Det col is
editable
• for Model 700/2350A, Mode 3, the Det col is
“DUAL” and read-only
CD
CD2
CDT associated with this stream (see
Section 5.2)
CDT 2 For the Model 700/2350A Modes 2 and
3, an additional CD column displays.
TE
TE2
TEV associated with this stream (see
Section 5.3)
TEV 2 For the Model 700/2350A Modes 2 and
3, an additional TE column displays.