D-6 ANALOG OUTPUT CAL. FOR 2350A
MON2000
Calibrating by Percentages JULY 2010
To calibrate by percentages,
1. Use the Application > Analog Outputs menu
to access this function.
2. The Analog Outputs dialog appears.
3. Select the channel to be calibrated.
Output channels 1 and 2 can be scaled from
either 4-20 mA or 1-5 V. Channels 3 to 10
are scaled from 4-20 mA.
All can be scaled with a percentage.
Although output channels for direct voltage
measurement exist for analog output channels 1 and 2,
it is recommended that they not be used for calibration
purposes. Instead, use the current-out terminals and
the 250-ohm resistor as described.