UPGRADE 2350A GC S/W AND 2350 EPROMS E-1
MON2000
JULY 2010 Connect to GC and Halt Analysis
EA
APPENDIX E, UPGRADE 2350A GC S/W AND2350 EPROMS
E.1 CONNECT TO GC AND HALT ANALYSIS
1. Ensure that the MON2000 software is
running and you are connected to the GC
that is to be upgraded.
Use the File > Connect menu to access this
function. See Section 2.10.1 for more
information.
2. Use MON2000 to halt any ongoing analysis
or calibration runs.
Use the Control > Halt menu to access this
function. See Section 4.3 for more
information.
E.2 OFFLINE EDIT TO UPLOAD APP. & RENAME
1. Use MON2000 to upload (from the GC unit)
to the PC the most current parameters that
are in use at the GC.
Use the File > Offline Edit > Upload
Application menu to access this function.
2. A menu of stored application files appears.
Double-click the desired file.
To upgrade the Model 500 2350A (GC) software and the
2350 EPROMs, perform all steps in the order they appear
in this appendix.
Before beginning the upgrade procedure, ensure that you
have identified which standard (i.e., factory-released)
application file was used to form the current user
application for this GC. See Section E.10 for more
information on standard applications.
When halted GC runs
have finished, the
status for Detector 1
mode will indicate Idle
(see GC Status bar; see
Section 3.3).