APPLICATION FUNCTIONS 5-65
MON2000
JULY 2010 FID Configuration
3. If Flame status is OFF, due to a power
failure or maintenance, for example, you
can restart the flame by doing the following:
(a) Click H2 Open. The H2 valve status
parameter displayed in the FID PreAmp
Status table changes to OPEN.
(b) Click Light flame. The Flame status
parameter displayed in the FID PreAmp
Status table will change to ON when the
FID internal temperature exceeds the
value set in the Flame On sense
temperature field.
If FID ignition is set to Auto, the GC will
automatically restart the flame if it goes
out.
4. To enable or disable digital filtering, do the
following:
(a) Right-click on the FID Configuration
window. The FID Advanced
Configuration window appears.