APPLICATION FUNCTIONS 5-41
MON2000
JULY 2010 Strea ms
For calibration (CAL) streams, you can edit the following control parameters:
Table 5-5 Description of Calibration Stream Settings
Cal Setting Description
Tot number of runs made for each calibration
Range: 1 to 10
Avg number of last calibration runs to
average
E.g., if 5 calibration runs are performed
and Avg is set to “3”, then the last 3 runs
are used to average the calibration
results.
Start time the first automatic calibration is to
be performed
Int number of hours between automatic
calibrations