D-2 ANALOG OUTPUT CAL. FOR 2350A
MON2000
Calibrating by Volts JULY 2010
3. Select the channel to be calibrated.
Output channels 1 and 2 can be scaled from
either 4-20 mA or 1-5 V. Channels 3 to 10
are scaled from 4-20 mA.
Although output channels for direct voltage measurement
exist for analog output channels 1 and 2, it is recommended
that they not be used for calibration purposes. Instead, use
the current-out terminals and the 250-ohm resistor as
described.