APPLICATION FUNCTIONS 5-37
MON2000
JULY 2010 Delayed Discrete Alarms
7. Select Applications > User Defined >
Numeric.... The User-Defined Numerics
window displays.
8. The following variables apply to the delayed
discrete alarm:
delay_alm_da: Defines which discrete
alarm will be activated.
delay_alm_di: Defines which discrete
input will be used to initiate the delayed
discrete alarm.
delay_time: Defines the time, in
seconds, between the moment the
discrete input is closed and the delayed
discrete alarm is activated. Range is 0 to
500 seconds. Default value is 60
seconds.
Although there are five discrete inputs available, closing
Discrete Input #4 will result in an active Analyzer Failure
alarm that will cause the GC to go to idle if Discrete
Input #4 remains closed up to the end of the current
analysis.
Discrete Input #5 can be used if the run_Aux_Seq option
in the 'User-Defined Selections window is set to OFF.