ANALOG OUTPUT CAL. FOR 2350A D-7
MON2000
JULY 2010 Calibrating by Percentages
4. To change a variable assignment, click the
appropriate Variable cell. Use the provided
pull-down menu and click the desired
variable to select it.
5. Set Zero Scale to “0” and Full Scale to
“1200”.
6. Set the Fixed/Var parameter to “Fixed”.
7. Set Fixed Value to “0.0”.
8. Set Zero Scale Adjustment and Full Scale
Adjustment to “0.0”.
9. Click the button to accept your
changes and exit from the Analog Outputs
dialog.
10.Connect a 250-ohm resistor across the
signal at the input of the device where this
analog output is used. Connect a Voltmeter
across the resistor and measure the result.
11.Return to the Analog Outputs dialog and
reselect the channel being calibrated.
12.Set the Fixed Value to the Full Scale value
entered in Step 5.
Your Voltmeter reading
should be close to 1.0 V
but does not have to be
exactly 1.0 V. If the
reading is negative,
reverse the leads.