INSTALLATION AND SETUP 2-45
MON2000
JULY 2010 Conversion Process
10.Remove the DSPI/O board assembly
(drawing P/N CE-12976) and cables from
the third slot of the card cage assembly.
This board is not used on the 2350A GC
Controller.
11.Remove any other optional boards, such as
memory expansion boards and associated
cables. These parts are not used on the
2350A GC Controller.
12.Leave the Analog Board assembly (drawing
P/N BE-18044) with attached cable in the
card cage. This board is used on the 2350A
basic configuration.
13.Change the fuse in the in-line fuse holder
from 1 amp to the 2.5 amp, Slo-Blo fuse
provided. The fuse is located in the cable (P/
N 2-3-2350-069) between the power supply
and the System Interface Board.
14.Ensure that DIP Switches 6, 7, and 8 of S1
are all in the "OFF" position. If the 2350A is
powered up with S8 in the OFF position, it
will delete the current application
(commonly called a "Cold Start").
15.To inspect or change the GC Controller's
COM ID setup at the GC Controller site,
locate the DIP switch as described in the
following steps.
16.For the explosion-proof Controller, the front
panel is secured by 16 screws. Remove those
screws first.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Before removing the unit cover from the GC Controller, make certain the power
supply switch is OFF and the AC power cord is disconnected. Observe all safety
precautions when you are working in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.