APPLICATION FUNCTIONS 5-13
MON2000
JULY 2010 View Raw Data
RT is automatically targeted by
MON2000). Then, press the DOWN
ARROW key to advance to the next 'Raw
Data' value.
(d) Repeat Step (c) until all Retention Times
have been copied to the CDT.
6. Click the button to clear the Raw
Data screen.
Following is an example of a Raw Data
Table.
Table 5-3 Raw Data Table
Description Method
Baseline Resolved 1
Fused 2
Last of Fused Group 3
Tangent Skimmed 4
Resolved Rider Peak 5
Fused Rider 6
Last of Fused Rider 7
Tailing Peak 8
Forward Horizontal 20
Backward Horizontal 40
Baseline Forced at Valley 60
Inhibit On 100
Forced Integrate On 200
Forced Integrate End 300
Plateau Detect 400
Summation 500
Some methods are a
combination of the above
types, e.g. 103 is LAST OF
FUSED GROUP (3) and
INHIBIT ON (100).