APPLICATION FUNCTIONS 5-9
MON2000
JULY 2010 Component Data
Fixed/Variable whether a component response factor is fixed or
variable
A fixed response factor will not be updated during
calibration.
Calibration Concentration component concentration in calibration gas
To enter a PPM, type “Xp”; e.g., “4.0p” will appear
as “4.0 ppm”. Values above 90 PPM are converted
to percentages.
Analysis Method used to determine component raw data value
Use the pull-down menu to select either area,
height, fixed, or the appropriate analog input
(used with auxiliary analyzer).
RT Percent Dev. maximum acceptable percentage of deviation of
the new retention time from the current retention
time
RT Secs. Dev. maximum acceptable deviation time, in seconds, of
the new retention time from the current retention
time
RT Update Method determines when component retention time will be
updated
Select Analysis to update after each run.
Select Calib to update only during final calibration
run.
Resp. Factor % Dev. maximum acceptable percent deviation between
the new response factor and the current response
factor
Total Concentration total concentration of all components
Table 5-2 Description of Standard Parameters in a CDT (Continued)