Hazardous Area Oxymitter 4000

Instruction Manual

IB-106-340C Rev. 4.1 July 2004

calibration). If the slope or the

9-3 CALIBRATION – HAZARDOUS AREA

constant is out of specifica-

OXYMITTER 4000 WITH LOI

tion, a diagnostic alarm LED

 

 

will be flashing. The diagnos-

Refer to Figure 6-4for a view of the LOI menu

tic alarm will remain active

tree. To calibrate the Hazardous Area Oxymitter

until the purge cycle is over. If

4000 from the LOI, access the CALIBRATION/

the three pattern flash occurs

Start Calibration menu.

without a diagnostic alarm, the

 

 

calibration gases could be the

a. CALIBRATION/ Start Calibration

same or the calibration gas

This is the starting point for calibrations. The

was not turned on.

LOI will instruct the user through this entire pro-

 

A flashing CAL LED indicates

cedure. You can select “Abort Calib” at any time

to abort the calibration.

the calibration is done. (See

Section 8, TROUBLE-

1. The LOI displays the following:

SHOOTING, for an explana-

tion of the 2 pattern and 3

Apply Gas 1

pattern flashes).

Hit E when ready

7 Remove the second calibra-

The Oxymitter 4000 is ready to accept

tion gas and cap off the cali-

the first calibration gas. Apply the first

bration gas port.

calibration gas. (Electronics will abort

8 Push the CAL key; the CAL

the calibration if this step is not done

within 30 minutes).

LED will be on solid as the

 

 

unit purges. (Default purge

2. Touch the Enter key to start the Gas 1

time is three minutes). When

flow. A timer is activated to allow the

the purge is complete, the

calibration gas adequate time to flow

CAL LED will turn off and the

(default time of five minutes). The LOI

Hazardous Area Oxymitter

displays:

4000 output unlocks from its

Flow Gas 1xxxxs

held value and begins to read

Read Gas 1xxxxs

the process O2.

Done Gas 1

If the calibration was valid, the

The display counts down the seconds

DIAGNOSTIC ALARMS LEDs will

remaining to flow Gas 1, then the time

indicate normal operation. If either

remaining for sensing the O2 concen-

new calibration value (slope or

tration of Gas 1. Done Gas 1 indicates

constant) is not within parameters,

completion.

the DIAGNOSTIC ALARMS LED

 

 

will indicate an alarm. (See Section

3. Remove the first calibration gas and

8, TROUBLESHOOTING, for alarm

apply the second calibration gas.

codes). If the calibration was inva-

(Electronics will abort the calibration

lid, the Hazardous Area Oxymitter

if this step is not done within 30 min-

4000 will return to normal opera-

utes). The LOI displays the following:

 

 

tion, as it was before a calibration

 

Apply Gas 2

 

was initiated, and the parameters

 

Hit E when ready

 

will not be updated.

 

9

(e) Place control loop in automatic.

 

 

 

 

 

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 9-5

Page 123
Image 123
Emerson Process Management OXYMITTER 4000 instruction manual Calibration Hazardous Area, Oxymitter 4000 with LOI