Hazardous Area Oxymitter 4000
Instruction Manual
calibration). If the slope or the |
| ||
constant is out of specifica- | OXYMITTER 4000 WITH LOI | ||
tion, a diagnostic alarm LED |
|
| |
will be flashing. The diagnos- | Refer to Figure | ||
tic alarm will remain active | tree. To calibrate the Hazardous Area Oxymitter | ||
until the purge cycle is over. If | 4000 from the LOI, access the CALIBRATION/ | ||
the three pattern flash occurs | Start Calibration menu. | ||
without a diagnostic alarm, the |
|
| |
calibration gases could be the | a. CALIBRATION/ Start Calibration | ||
same or the calibration gas | This is the starting point for calibrations. The | ||
was not turned on. | |||
LOI will instruct the user through this entire pro- | |||
| |||
A flashing CAL LED indicates | cedure. You can select “Abort Calib” at any time | ||
to abort the calibration. | |||
the calibration is done. (See | |||
Section 8, TROUBLE- | 1. The LOI displays the following: | ||
SHOOTING, for an explana- | |||
tion of the 2 pattern and 3 | Apply Gas 1 | ||
pattern flashes). | Hit E when ready | ||
7 Remove the second calibra- | The Oxymitter 4000 is ready to accept | ||
tion gas and cap off the cali- | the first calibration gas. Apply the first | ||
bration gas port. | calibration gas. (Electronics will abort | ||
8 Push the CAL key; the CAL | the calibration if this step is not done | ||
within 30 minutes). | |||
LED will be on solid as the |
|
| |
unit purges. (Default purge | 2. Touch the Enter key to start the Gas 1 | ||
time is three minutes). When | flow. A timer is activated to allow the | ||
the purge is complete, the | calibration gas adequate time to flow | ||
CAL LED will turn off and the | (default time of five minutes). The LOI | ||
Hazardous Area Oxymitter | displays: | ||
4000 output unlocks from its | Flow Gas 1xxxxs | ||
held value and begins to read | Read Gas 1xxxxs | ||
the process O2. | Done Gas 1 | ||
If the calibration was valid, the | The display counts down the seconds | ||
DIAGNOSTIC ALARMS LEDs will | remaining to flow Gas 1, then the time | ||
indicate normal operation. If either | remaining for sensing the O2 concen- | ||
new calibration value (slope or | tration of Gas 1. Done Gas 1 indicates | ||
constant) is not within parameters, | completion. | ||
the DIAGNOSTIC ALARMS LED |
|
| |
will indicate an alarm. (See Section | 3. Remove the first calibration gas and | ||
8, TROUBLESHOOTING, for alarm | |||
apply the second calibration gas. | |||
codes). If the calibration was inva- | |||
(Electronics will abort the calibration | |||
lid, the Hazardous Area Oxymitter | |||
if this step is not done within 30 min- | |||
4000 will return to normal opera- | |||
utes). The LOI displays the following: |
| ||
| |||
tion, as it was before a calibration |
| ||
Apply Gas 2 |
| ||
was initiated, and the parameters |
| ||
Hit E when ready |
| ||
will not be updated. |
| 9 | |
(e) Place control loop in automatic. |
|
| |
|
|
|
Rosemount Analytical Inc. A Division of Emerson Process Management | Maintenance and Service |