Instruction Manual

IB-106-340C Rev. 4.1 July 2004

Hazardous Area Oxymitter 4000

OXYGEN

SIGNAL

PROBE

REMOTE

ELECTRONICS

4-20 mA

SIGNAL

INTERCONNECTING

 

 

 

WIRING DIAGRAM

 

 

CABLE

 

 

 

TO ELECTRONICS

 

 

 

 

 

 

 

 

 

GRN

ORN

YEL

RED

 

 

BLK

BLK

 

1

2

3

4

5

6

7

8

 

 

O2

 

 

T/C

 

 

HTR

 

 

CELL

ORN

YEL

RED

 

 

WHT

BLK

GENERAL PURPOSE

BLU

 

G.P.

 

OXYMITTER

(WHT)

 

(BRN)

(RED)

(GRN)

(HAZ.)

 

(BLU)

(BRN)

HAZARDOUS AREA

 

 

OXYMITTER

 

 

 

 

 

 

 

 

 

FROM PROBE

TO IMPS 4000 IF USED.

LOGIC I/O +

 

REFER TO IMPS 4000

 

LOGIC I/O -

 

INSTRUCTION MANUAL.

 

 

 

 

4-20 mA +

 

 

4-20 mA -

 

 

GROUND

 

90-250 VAC, 50-60 HZ

LINE 1

NEUTRAL

LINE VOLTAGE INPUT

 

 

GROUND

 

 

 

 

 

 

TYPE K

 

 

 

 

 

 

THERMOCOUPLE

 

 

 

 

 

 

SIGNAL

 

 

 

 

 

 

 

 

 

HEATER POWER

 

 

 

 

 

 

 

 

 

 

-WHT

+

 

GRN-

+

BLU BRN

 

(BELOW COVER)

 

 

 

 

 

 

BRN

 

 

RED

 

 

 

 

 

 

 

 

 

 

 

 

 

TERMINAL

GRN/YEL

BLOCK

or GRN

 

GROUND

LUGS

CALIBRATION

HANDSHAKE/

LOGIC I/O

LINE VOLTAGE

+ 4-20 -

 

 

 

 

-

 

+

AC

AC

N

L1

TERMINAL

BLOCK

GROUND LUGS

37270023

Figure 2-10. Electrical Installation – Hazardous Area Oxymitter 4000

with Remote Electronics (Sheet 1 of 2)

2-12 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 58
Image 58
Emerson Process Management OXYMITTER 4000 instruction manual Wiring Diagram