Instruction Manual

IB-106-340C Rev. 4.1 July 2004

Hazardous Area Oxymitter 4000

(CONTINUED FROM

SHEET 1)

SYSTEM

Calib Setup

Input/Output

Parameters

Software

Status

Analog

Digital

O2 Gas 1

_____%

O2 Gas 2

_____%

O2-Reset Vals

Yes/No

O2 Out Tracks

Yes/No

O2 Cal Intervl

____H

O2-Next Cal

____H

Gas Time

___Sec.

Purge Time

___Sec.

Auto Calib?

Yes/No

 

 

 

 

O2 Type

_______

O2 Range

______%

O2 Alarm Level

_____mA

Do O2 Trim

 

 

 

 

 

Logic IO Mode

See Table 4-1

Low O2 Alarm See para. 9-3b

 

Input State

_______

 

Force Output

_______

 

 

 

 

 

 

 

O2 Slope

____ mV/D

 

O2 Constant

____ mV

 

O2 T90 Time

0:00

 

Auto Tune?

Yes/No

 

Lockout Time

0:00

 

Revert Time

0:00

 

Luminance

______

 

 

 

 

 

 

 

Version

xxx

 

Checksum

xxx

 

Build Number

xxx

 

Build Date

xxxxxx

 

Test Code

xx

 

SW Err File

xx

 

SW Err Line

xx

 

 

 

 

 

 

 

Alarms

__________

 

PID Parameters

115/220

 

Reset Device?

Yes/No

(Cal. required after reset)

NOTE

In column four of this menu, the selections in Italics are user configurable. Bold text selections are procedures; related instructions are displayed on the LOI. All other parameters are display only.

37260018

Figure 6-4. Local Operator Interface Menu Tree (Sheet 2 of 2)

6-6 HAZARDOUS AREA OXYMITTER 4000 SETUP AT THE LOI

In setting up the Hazardous Area Oxymitter 4000 from the LOI, it’s best to start at the SYSTEM/Calibration Setup menu, Figure 6-4.

a.SYSTEM/Calibration Setup

1.O2 Gas #1 – Enter the high or low cal gas value (the order is not important).

2.O2 Gas #2 – Enter the second cal gas value.

NOTE

NOTE

Rosemount Analytical recommends 0.4% O2 and 8% O2 for calibration gases.

3.O2 - Reset Values – Resets factory default values.

4.O2 Output Tracks – 4 to 20 mA signal can be held at the last value during calibration, or the signal can be left to track the cal gases.

5.O2 Cal Interval – If automatic calibra- tion is selected, this selects the interval between calibrations.

Refer to Section 9, MAINTENANCE AND SERVICE, for calibration instructions.

6.O2-Next Cal – If automatic calibration is selected, this selects the time until the first initial calibration takes place.

6-4 Startup and Operation with LOI

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 82
Image 82
Emerson Process Management OXYMITTER 4000 Hazardous Area Oxymitter 4000 Setup AT the LOI, SYSTEM/Calibration Setup