Instruction Manual
IB-106-340C Rev. 4.1
July 2004
6-4 Startup and Operation with LOI Rosemount Analytical Inc. A Division of Emerson Process Management
Hazardous Area Oxymitter 4000
SYSTEM
O2 Gas 1
O2 Gas 2
O2-Reset Vals
O2
O2 Cal Intervl
O2-
Gas Time
Purge Time
_____%
_____%
____H
___Sec.
Yes/No
Yes/No
____H
___Sec.
Yes/No
Out Tracks
Next Cal
Auto Calib?
O2 Type
O2 Range
O2 Alarm Leve
_______
______%
l _____mA
Do O2 Trim
Calib Setup
Input/Output
Parameters
Status
Software
37260018
(CONTINUED FROM
SHEET 1)
Analog
Digital
NOTE
In column four of this menu, the selections in are user configurable. text selections are
procedures; related instructions are displayed on the LOI. All other parameters are display only.
Italics Bold
Version xxx
Checksum xxx
Build Number xxx
Build Date xxxxxx
TestCode xx
SW Err File xx
SW Err Line xx
Alarms __________
PID Parameters 115/220
Yes/No
(Cal. required after reset)
Reset Device?
O2 Slope
O2 Constant
O2 T90 Time
Auto Tune?
Lockout Time
Revert Time
Luminance
____ mV/D
____ mV
0:00
Yes/No
0:00
0:00
______
Logic IO Mode
Low
See Table4-1
O2 Alarm
Force Output
See para. 9-3b
Input State _______
_______
Figure 6-4. Local Operator Interface M enu Tree (Sheet 2 of 2)
6-6 HAZARDOUS AREA OXYMITTER 4000
SETUP AT THE LOI
In setting up the Hazardous Area Oxymitter
4000 from the LOI, it’s best to start at the
SYSTEM/Calibration Setup menu, Figure 6-4.
a. SYSTEM/Calibration Setup
1. O 2 Gas #1 – Enter the high or low cal
gas value (the order is not important).
2. O2 G as #2 – Enter the second cal gas
value.
NOTE
Refer to Section 9, MAINT ENANCE AND
SERVICE, for calibration instructio ns.
NOTE
Rosemount Analytical recommends
0.4% O2 and 8% O2 for calibration gases.
3. O2 - Res et Values – Resets factory
default values.
4. O2 O utput Tracks – 4 to 20 mA signal
can be held at the last value during
calibration, or the signal can be left to
track the cal gases.
5. O 2 Cal Interval – If automatic calibra-
tion is selected, this selects the interval
between calibrations.
6. O2- Next Cal – If automatic calibration
is selected, this selects the time until
the first initial calibration takes place.