AlarmsVery Hi O2 Temp
Figure 8-8. Fault 6, High High Heater Temp
f. Fault 6, High High Heater Temp
Figure 8-8 shows the electronic assembly
for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).
Membrane Keypad. When Fault 6 is
detected, the HEATER LED flashes twice,
pauses for three seconds, and repeats.
1. T he high high heater temp alarm will
activate when the thermocouple pro-
duces a voltage of 37.1 mV
[900°C (1652°F)].
2. T he triac and the temperature control
may be at fault.
3. Rem ove power. Allow Hazardous Area
Oxymitter 4000 to cool for five minutes.
Restore power.
4. If the condition repeats, replace the
electronic assembly per paragraph
9-4c.1, Electronic Assembly
Replacement.
LOI. When Fault 6 is detected, the LOI
displays the “Very Hi O2 Temp” message.
1. T he very high O2 temperature alarm
will activate when the thermocouple
produces a voltage of 37.1 mV
[900°C (1652°F)].
2. T he triac and the temperature control
may be at fault.
3. Rem ove power. Allow the Oxymitter
4000 to cool for five minutes. Restore
power.
4. If the condition repeats, replace the
electronic assembly per paragraph
9-4c, Electronic Assembly Replace-
ment.