Hazardous Area Oxymitter 4000

Instruction Manual

IB-106-340C Rev. 4.1 July 2004

4.CAL LED. The CAL LED is on steady or flashing during calibration. Further information is available in Section 9,

MAINTENANCE AND SERVICE.

5.Keys.

(a)INC and DEC. The INC and DEC keys are used to set the values of the calibration gases. Attach a multimeter across TP5 and TP6. The calibration and process gases can now be monitored. Pressing the INC or DEC once will cause the output to switch from the proc- ess gas to the calibration gas. Pressing INC or DEC a second time will increase or decrease the calibration gas parameter. If the keys have been inactive for one minute, the output reverts to the process gas. When a calibration has been initiated, the value at TP5 and TP6 is the % O2 seen by the cell.

Oxygen levels, as seen on the multimeter, are:

8.0% O2 = 8.0 volts DC

0.4% O2 = 0.4 volts DC

(b) CAL. The CAL key can:

 

1

Initiate a calibration.

 

2

Sequence through calibration.

 

3

Abort the calibration.

 

 

NOTE

 

Refer to Section 9, MAINTENANCE AND

 

SERVICE, for calibration instructions.

 

b. Model 751 Remote Powered Loop

 

 

LCD Display (Optional)

 

Refer to Remote Powered Loop LCD

5

manual for calibration and operation.

 

 

 

 

HEATER T/C

DIAGNOSTIC HEATER

ALARMS O2 CELL CALIBRATION

CALIBRATION RECOMMENDED

SW2

ON

HEATER T/C

HEATER

O2 CELL

CALIBRATION

TEST POINTS

INC INC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TP1

J1

 

TP2

 

TP3

 

TP4

RED YEL GRN ORG

 

1

2

3

4

1

2

3

4

LIGHTING SEQUENCE DURING NORMAL OPERATION

HIGH LOW

GAS GAS

DEC DEC

TEST GAS + PROCESS - % O2

CAL LED

TP5

TP6

22220055

Figure 5-3. Normal Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation with Keypad 5-3

Page 77
Image 77
Emerson Process Management OXYMITTER 4000 instruction manual Normal Operation