Instruction Manual
IB-106-340C Rev. 4.1
July 2004
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation with Keypad 5-1
Hazardous Area Oxymitter 4000
SECTION 5START UP AND OPERATIONOF HAZARDOUS AREA OXYMITTER 4000WITH MEMBRANE KEYPAD
5-1 POWER UP
a. Startup Display
When power is applied to the probe, the cell
heater turns on. It takes approximately one
half hour for the cell to heat to operating
temperature. This condition is indicated b y
the top four LEDs (DIAGNOSTIC ALARMS)
on the membrane keypad (Figure 5-1).
Starting with the CALIBRATION LED, the
LEDs light in ascending order until all four
LEDs are on. At this point, all four turn off
and the cycle starts again. This ramp cycle
continues until the cell is up to operating
temperature.
b. Operating Display
The ramp cycle turns into a cycle where the
diagnostic LEDs light in sequence from the
top to the bottom, one at a time. After the
bottom LED turns on, the sequence starts
again at the top with the HEATER T/C LED
(Figure 5-1).
c. Error
If there is an error condition at startup, one
of the diagnostics LEDs will be blinking.
Refer to Section 8 TROUBLESHOOTING,
to determine the cause of the error. Clear
the error, cycle power, and the operating
display should return.
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATERT/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATERT/C +
HEATERT/C -
INC INC
DEC DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1 J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
HEATERT/C
HEATER
O CELL
2
CALIBRATION
LIGHTING SEQUENCE DURING NORMAL OPERATION
(OPERATINGDISPLAY)
12 3 4 1 2 3 4
HEATERT/C
HEATER
O CELL
2
CALIBRATION
LIGHTING SEQUENCE DURING WARM-UP
(STARTUPDISPLAY)
12 3 4 1 2 3 4
22220056
Figure 5-1. Startup and Normal Operatio n
5