Emerson Process Management OXYMITTER 4000 Alarms O2 Heater Open, Fault 5, Open Heater

Models: OXYMITTER 4000

1 150
Download 150 pages 29.63 Kb
Page 105
Image 105

Hazardous Area Oxymitter 4000

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

e. Fault 5, Open Heater

Figure 8-7shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with

a membrane keypad (upper view) and a

Hazardous Area Oxymitter 4000 with an

LOI (lower view).

Membrane Keypad. When Fault 5 is

TEST

POINTS

INC INC

HIGH LOW

GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

detected, the HEATER LED flashes once,

pauses for three seconds, and repeats.

1.

Remove power.

2.

Remove the electronic assembly per

 

paragraph 9-4c,Electronic Assembly

 

Replacement.

3.

Using a multimeter, measure the re-

 

sistance across the terminals of heater

 

connector, J8.

KEYPAD

Alarms

O2 Heater Open

LOI

37260023

Figure 8-7. Fault 5, Open Heater

4. The measurement should be approxi-

mately 72 ohms. If the heater is open,

see paragraph 9-4g,Heater Strut

Replacement.

LOI. When Fault 5 is detected, the LOI dis- plays the “O2 Heater Open” message.

1.Remove power. Unscrew and remove the LOI module from the electronic as- sembly.

2.Perform the diagnostic steps 2 through 4 shown for the membrane keypad.

8

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-9

Page 105
Image 105
Emerson Process Management OXYMITTER 4000 Alarms O2 Heater Open, Fault 5, Open Heater, Membrane Keypad. When Fault 5 is