Hazardous Area Oxymitter 4000

Instruction Manual

IB-106-340C Rev. 4.1 July 2004

8-4 ALARM CONTACTS

2. Additional IMPS 4000 Alarm Contacts.

a. If autocalibration is not utilized, a common

(a)

One contact per IMPS 4000 for

bi-directional logic contact is provided for

 

“low calibration gas flowing”.

any of the diagnostic alarms listed in Table

 

 

8-1.The assignment of alarms which can

(b)

One contact per IMPS 4000 for

actuate this contact can be modified to one

 

“high calibration gas flowing”.

of seven additional groupings (mode 0

 

 

through mode 7) listed in Table 7-1.

NOTE

 

The logic contact is self-powered, +5 VDC,

The 4-20 mA signal can be configured

 

with a 340 ohm series resistance. An inter-

to respond normally during any cali-

 

posing relay will be required if this contact is

bration, or can be configured to hold

 

to be utilized to annunciate a higher voltage

the last O2 value upon the initiation of

 

device, such as a light or horn. An inter-

calibration. Factory default is for the

 

posing relay may also be required for cer-

4-20 mA signal to operate normally

 

tain DCS input cards.

throughout calibration.

 

A Potter & Brumfield R10S-E1Y1-J1.0K 3.2

NOTE

 

mA DC or an equal interposing relay will be

Holding the last O2 value may be useful

 

mounted where the contact wires terminate

 

in the control/relay room.

if several probes are being averaged

 

 

 

for the purpose of automatic control.

 

b. If autocalibration systems are utilized, the

Unless several probes are being aver-

 

bi-directional logic contact is utilized as a

aged, always place any control loops

 

“hand-shake” signal between the autocali-

using the O2 signal into manual prior to

 

bration system (SPS 4000 or IMPS 4000)

calibrating.

 

and is unavailable for alarming purposes.

 

 

The following additional contacts are pro-

8-5 IDENTIFYING AND CORRECTING ALARM

 

vided through the autocalibration systems:

 

1. SPS 4000 and IMPS 4000, 1-4 probes.

INDICATIONS

 

 

 

(a)

One contact closure per probe

For a Hazardous Area Oxymitter 4000 with a

 

membrane keypad, faults are indicated by four

 

 

from the control room to the SPS

 

 

diagnostic, or unit, alarm LEDs. A pattern of

 

 

4000 or IMPS 4000 for “calibration

 

 

repeating blinks define the problem. A con-

 

 

initiate”.

 

 

densed table of the errors and the correspond-

 

(b)

One contact output per probe from

ing blink codes can be found on the inside right

 

cover of the electronics housing. Table 8-1also

 

 

 

the SPS 4000 or IMPS 4000 to the

identifies the blink code and fault status of each

 

 

control room for “in calibration”

LED as well as the output of the 4-20 mA signal

 

 

notification.

line and a fault number that corresponds to the

8

(c) One contact output per probe from

troubleshooting instructions provided in this

 

section.

 

 

the SPS 4000 or IMPS 4000 to the

 

 

 

 

 

 

control room for “calibration failed”

For a Hazardous Area Oxymitter 4000 with the

 

 

notification. (Includes output from

optional LOI, alarm messages are displayed on

 

 

pressure switch indicating “cal gas

the LOI display window when the alarm status

 

 

bottles empty”).

display is accessed via the LOI menu. A listing

 

of the alarm/fault messages and the related fault status descriptions and fault numbers are shown in Table 8-2.

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-3

Page 99
Image 99
Emerson Process Management OXYMITTER 4000 instruction manual Alarm Contacts, Identifying and Correcting Alarm, Indications