i. Fault 9, High Heater Temp

Figure 8-11shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with

a membrane keypad (upper view) and a

Hazardous Area Oxymitter 4000 with an

LOI (lower view).

Membrane Keypad. When Fault 9 is

TEST

POINTS

 

 

 

INC

 

INC

 

 

 

HIGH

 

LOW

GAS

 

GAS

DEC

 

DEC

 

 

 

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5 TP6

J1 RED YEL GRN ORG

detected, the HEATER LED flashes five

times, pauses for three seconds, and

repeats.

1.

If the thermocouple produces a voltage

 

in excess of approximately 30.7 mV,

 

the high heater temp alarm activates.

2.

The 4-20 mA signal returns to the

 

default value (4 or 20 mA).

3.

This alarm is self-clearing. When tem-

 

perature control is restored and the

KEYPAD

Alarms

O2 Temp Hi

LOI

37260027

Figure 8-11. Fault 9, High Heater Temp

thermocouple voltage returns to the

normal range, the alarm clears.

4. If the temperature continues to rise, the

next alarm will be the high high heater

temp alarm.

LOI. When Fault 9 is detected, the LOI displays the “O2 Temp Hi” message. Refer to the comments and procedures in para- graphs 1 through 4 above.

8

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-13

Page 109
Image 109
Emerson Process Management OXYMITTER 4000 Alarms O2 Temp Hi, Fault 9, High Heater Temp, Membrane Keypad. When Fault 9 is