Figure 8-11. Fault 9, High Heater Temp
i. Fault 9, High Heater Temp
Figure 8-11 shows the electronic assembly
for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).
Membrane Keypad. When Fault 9 is
detected, the HEATER LED flashes five
times, pauses for three seconds, and
repeats.
1. If the thermocouple produces a voltage
in excess of approximately 30.7 mV,
the high heater temp alarm activates.
2. T he 4-20 mA signal returns to the
default value (4 or 20 mA).
3. T his alarm is self-clearing. When tem-
perature control is restored and the
thermocouple voltage returns to the
normal range, the alarm clears.
4. If the temperature continues to rise, the
next alarm will be the high high heater
temp alarm.
LOI. When Fault 9 is detected, the LOI
displays the “O2 Temp Hi” message. Refer
to the comments and procedures in para-
graphs 1 through 4 above.