Hazardous Area Oxymitter 4000

Instruction Manual

IB-106-340C Rev. 4.1 July 2004

SECTION 3

CONFIGURATION OF HAZARDOUS AREA OXYMITTER 4000 WITH MEMBRANE KEYPAD

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.

3-1 GENERAL

a.Verify Mechanical Installation

Ensure the Hazardous Area Oxymitter 4000 is installed correctly. See Section 2,

INSTALLATION.

Opening the electronics housing in hazardous areas may cause an explo- sion causing severe injury, or death. It may be required to get a hot work permit from your company safety offi- cer before you open the housing.

b. Verify Terminal Block Wiring

 

1. Remove screw (18, Figure 9-3or

 

Figure 9-4), cover lock (19), and cap-

3

tive washer (20) that secure cover (17)

on left side of housing (11). Remove the cover.

2. Check the terminal block wiring (Figure 3-1). Be sure the power, the

4-20 mA signal, and the logic outputs are properly connected and secure. To avoid a shock hazard, the power terminal cover must be installed.

For units with remote electronics, check the terminal block wiring at the probe and at the remote electronics unit.

3. Install the cover (17, Figure 9-3or Figure 9-4)over terminal block (15) and secure with captive washer (20), cover lock (19), and screw (18).

+ 4-20 -

-

+

 

 

 

AC

AC

 

 

N

L1

4-20 mA

 

SIGNAL

TERMINAL

 

BLOCK

LOGIC I/O

 

Rosemount Analytical Inc.

SMART FAMILY

Orrville, OH 44667-0901

800-433-6076

HART

OXYMITTER 4000

SERIAL NO.

TAG NO.

VOLTS: 85-264 VAC

48-62 Hz WATTS:

OUTPUT: 4-20 mA

LINE FUSE:

HAZARDOUS AREA

OXYMITTER 4000

ELECTRONICS

HOUSING

500 VA

5 Amps

HEATER T/C

DIAGNOSTIC HEATER SW2

ALARMS 02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

 

 

02 CELL mV +

TP1

J1

TEST

02 CELL mv -

TP2

 

POINTS

HEATER T/C +

TP3

 

 

 

HEATER T/C -

TP4

RED YEL GRN ORG

 

 

 

 

INC

INC

CAL

 

 

 

 

 

 

HIGH

LOW

 

 

 

GAS

GAS

 

 

 

DEC

DEC

TEST GAS +

TP5

 

PROCESS -

TP6

 

 

 

 

 

 

% 02

 

 

GROUND LUGS

37270002

Figure 3-1. Electronics Housing Terminals and Membrane Keypad

Rosemount Analytical Inc. A Division of Emerson Process Management

Configuration with Keypad 3-1

Page 63
Image 63
Emerson Process Management OXYMITTER 4000 instruction manual General, Verify Mechanical Installation