4. REMOTE OPERATION

GPIB

Purpose

Read or set the GPIB interface address.

Command

“GPIB addr

 

 

Query

“GPIB?”

 

 

Classic

“GPIB=addr

 

 

 

“GPIB”

 

 

Defaults

“GPIB 10”

 

 

Arguments

Addr:

The address of the IEEE-488 (GPIB) interface (1 to 30)

Remarks

The GPIB address is changed following the reply of this command. Each device on a GPIB

 

interface bus requires a unique address.

Example

Cmd sent:

“GPIB 21”

 

(enhanced)

Query reply

“21”

(No reply if GPIB-488)

Example

Cmd sent:

“GPIB? 21”

 

(enhanced)

Query reply

“21”

 

Example

Sent:

“GPIB=21”

 

(classic)

Reply:

“21”

 

Errors

ERR# 6:

The argument is not within given limits.

See Also:

3.5.2.2

 

 

 

 

 

 

HEAD

 

 

 

Purpose

Read or set the fluid head settings.

 

Command

“HEAD height, units, fluid

 

Query

“HEAD?”

 

 

Classic

“HEAD=height, units, fluid

 

 

“HEAD”

 

 

Defaults

“HEAD 0, cm, N2”

 

Arguments

height:

The height of the test in relation to the RPM4. The value is positive if the test is

 

 

above the RPM4, or negative if below the RPM4. The value can be between -

 

 

9999 and 9999. Setting the value to ‘0’ disables the head correction.

 

units:

The height units. This must be “in” or “cm”.

 

gas:

The fluid type. This must be “N2”, “Air”, “He”, “Oil”, “H2O”, or “User”.

Remarks

The RPM4 can make a fluid head correction to allow it to display the pressure at a level other than

 

the reference level of the RPM4.

 

Example

Cmd sent:

“HEAD? 10,in,N2”

 

(enhanced)

Query reply:

“10, in, N2”

 

Example

Sent:

“HEAD=10,in,N2”

 

(classic)

Reply:

“10, in, N2”

 

Errors

ERR# 2:

The fluid text is too long

 

ERR# 6:

The height is not within given limits or the unit is invalid.

 

ERR #7:

The fluid text is invalid

See Also:

3.3.7, 3.5.3

 

 

 

 

 

 

ID

 

 

 

Purpose

Read or set the user defined instrument identification label.

Command

“ID string

 

 

Query

“ID”

 

 

Classic

“ID=string

 

 

 

“ID”

 

 

Arguments

String:

An alphanumeric string up to 12 characters long.

Remarks

The user defined ID label can be used to allow the user to “tag” the RPM4 with a unique

 

identifier. This ID is stored in non-volatile memory and cannot be erased by a power failure,

 

system fault or reset. The ID should not be changed frequently, as the non-volatile memory

 

may fail after 100,000 write operations.

Example

Cmd sent:

“ID RPM4 #A01”

 

(enhanced)

Query reply

“RPM4 #A01”

(No reply if GPIB-488)

Example

Cmd sent:

“ID? RPM4 #A01”

 

(enhanced)

Query reply

“RPM4 #A01”

 

Example

Sent:

“ID=RPM4 #A01”

 

(classic)

Reply:

“RPM4 #A01”

 

Errors

ERR# 2

Text argument is too long (more than 12 characters)

See Also

3.5.5.4

 

 

Page 93