3. OPERATION
The AutoZ function can be turned ON and OFF. Table 7 summarizes the effect of AutoZ ON and OFF for the three
AutoZ in absolute measurement mode with an Axxx Q-RPT
In absolute measurement mode, the source of Pstd,0 must be an absolute pressure, nominally atmospheric pressure, with uncertainty significantly better than that of the
When using an Axxx (absolute)
Pabs = Pu - Poffset,A
In absolute measurement mode, running the AutoZ function to update the AutoZ value (Poffset,A), is initiated by the operator using [AutoZ].
AutoZ in gauge and negative gauge measurement modes with an Axxx
PRINCIPLE). Gauge measurement mode is achieved by subtracting the value of atmospheric pressure from the
In gauge measurement modes on an Axxx
Pgauge = Pu - Poffset,G
When AutoZ is ON, dynamic compensation for atmospheric pressure is also applied to compensate for changes in atmospheric pressure as measured by a barometer (ΔPatm) between AutoZ updates (see Section 3.2.2). The measured gauge pressure is calculated using ΔPatm to correct the value of Poffset.G.
Pgauge = Pu - Poffset.G - ΔPatm
In gauge measurement mode on an Axxx RPT, running the AutoZ function to update the value of Patm,0, is initiated by the operator using [AutoZ]. Updating of ΔPatm occurs automatically and continuously.
When gauge or negative gauge mode is first activated on an Axxx
Poffset.G subtracted from the absolute pressure reading to achieve gauge pressure is standard atmosphere of 101.325 kPa. This can cause a large zero error when gauge mode
is activated if atmospheric pressure is significantly different from standard atmospheric pressure. When gauge mode is activated, always press [AutoZ] to run AutoZ and obtain an actual value of Poffset.G .