RPM4™ OPERATION AND MAINTENANCE MANUAL

PRINCIPLE

This feature not available at the time of release of this manual. Check www.dhinstruments.com for new RPM4 embedded software releases and manual updates.

OPERATION

Feature not available at the time of release of this manual.

3.4[SETUP]

PURPOSE

[SETUP] accesses a menu of functions and features commonly used in setting up and using RPM4.

OPERATION

To access the SETUP menu, press [SETUP] from the MAIN RUN screen. The display is:

SETUP menu choices include:

1range 2res 3stab 4UL 5ATest

<1range>

Save and delete ranges created by AutoRange (see Section 3.4.1).

<2res>

Adjust the resolution of pressure displays (see Section 3.4.2).

<3stab>

Adjust the pressure stability test that is the criterion for the Ready/Not Ready indication

 

(see Section 3.3.3).

<4UL>

Adjust upper and lower pressure limit alarms (see Section 3.4.4).

<5ATest>

View data files resulting from running QDUT AutoTests (See Section 3.4.5).

 

3.4.1

<1RANGE>

PURPOSE

Save AutoRange ranges and associated settings for recall. Delete previously saved AutoRange ranges.

PRINCIPLE

The RPM4 AutoRange function creates a range and automatically sets various operating parameters proportionally to the range (see Section 3.3.4). Operating settings can then be changed by the user. Many operating settings are specific to the current range (see Table 4).

[SPECIAL], <1range> allows the AutoRange range and associated settings to be saved for recall using the [RANGE] key (see Section 3.3.1). This can avoid having to recreate frequently used ranges and settings.

[SPECIAL], <1range> is also used to delete ranges that have been saved but are no longer needed.

3.4.1.1SAVING AN AUTORANGE RANGE

OPERATION

To save a range created by AutoRange, the range must be the active range. Use AutoRange to create the range (see Section 3.3.4) and then make any desired feature and setting adjustments.