4. REMOTE OPERATION

This register can be read using the “*RSR?” query, Each of these status bits can set the RSB bit of the Status Byte Register, causing a SRQ to occur IF the RSB bit is enabled to do so. The Standard Event Status Enable Register (“*RSE” program message ) determines which of these flags are able to assert the RSB bit. The description of these bits are given as:

MEAS LO Lo-RPT measurement ready Bit 6 (64)

Indicates that the Hi Q-RPT has completed a Q-RPT measurement. NRDY LO Lo Q-RPT pressure Not Ready Bit 5 (32)

Indicates that the Hi Q-RPT made a transition from Ready to Not Ready as defined by the stability settings (see Section 3.2.1).

RDY LO Lo Q-RPT pressure Ready Bit 4 (16)

Indicates that the Hi Q-RPT is Ready as defined by the stability settings (see Section 3.2.1).

MEAS HI Hi Q-RPT measurement ready Bit 2 (4)

Indicates that the Hi Q-RPT has completed a Q-RPT measurement. NRDY HI Hi Q-RPT pressure Not Ready Bit 1 (2)

Indicates that the Hi Q-RPT made a transition from Ready to Not Ready as defined by the stability settings (see Section 3.2.1).

RDY HI Hi Q-RPT pressure Ready Bit 0 (1)

Indicates that the Hi Q-RPT is Ready as defined by the stability settings (see Section 3.2.1).

4.6IEEE STD. 488.2 COMMON AND STATUS PROGRAM MESSAGES

The RPM4 supports a set of commands that are common to all instruments conforming to IEEE Std. 488.2. These command make it easy to perform basic function for any device that supports these commands. These commands also cover the status reporting commands. See Section 4.5 for details on the status registers mentioned in these commands.

 

Table 21. Program Message List

 

 

CLS

Clear all of the status & event structures.

ESE

Read or set the Standard Event Status Enable Register.

ESR

Read the Standard Event Status Register.

IDN

Identify the RPM4 version, range, and serial number.

OPC

Set the operation complete bit when all operations have completed.

OPT

Read the list of installed RPM4 options.

RST

Reset the RPM4 settings to factory settings.

TST

Read the power on self test status.

SRE

Read or set the Service Request Enable Register.

STB

Read the Status Byte Register.

RSE

Read or set the Ready Status Enable Register.

RSR

Read the Ready Status Register.

Page 111

© 2004-2007 DH Instruments, a Fluke Company

Page 121