3. OPERATION

The AutoRange function simplifies the task of selecting the Q-RPT best suited to cover a specific test range and setting measurement parameters appropriately for the range. These are set automatically based on operator entry of desired measurement mode, pressure unit of measure and range full scale. The selections and settings made by AutoRange are summarized in Table 6.

A range set up using AutoRange and all its range specific settings, may be saved for recall using [SETUP], <1range> (see Section 3.4.1). Saved ranges are available for later selection using [RANGE] (see Section 3.3.1). If a range created by AutoRange is not saved, it is overwritten by the next AutoRange or erased when another range is selected.

Press [ENT] from the main run screen in RATE mode to use the [AutoTest] function to AutoRange RPM4 based on DUT tolerance (see Section 3.3.10).

Table 6. Settings Made by AutoRange

SETTING/SELECTION

PURPOSE

SET TO

SECTION

Unit

Set pressure unit of measure

Operator specified during

3.3.2

 

 

AutoRange

 

 

 

 

 

Mode

Set pressure measurement mode

Operator specified during

3.3.3

 

(absolute, gauge, negative gauge)

AutoRange

 

 

 

 

 

Full scale

Establish span maximum

Operator specified during

None

 

 

AutoRange

 

 

 

 

 

AutoRange span

Basis of span based setting

Absolute mode: Zero absolute

None

 

calculations

to AutoRange full scale

 

 

 

Gauge mode: Zero gauge to

 

 

 

AutoRange full scale

 

 

 

Negative gauge mode: Minus

 

 

 

atmosphere to AutoRange full

 

 

 

scale (negative equivalent of full

 

 

 

scale on BG15K Q-RPT).

 

 

 

 

 

Q-RPT

Select most appropriate Q-RPT

Absolute and negative gauge

1.2.2.1

 

to cover range and mode.

mode: Axxx Q-PRT with lowest

 

 

 

default full scale that is greater

 

 

 

than the AutoRange full scale

 

 

 

Gauge mode: Q-RPT with

 

 

 

lowest span whose full scale is

 

 

 

greater than AutoRange full

 

 

 

scale, Gxxx over Axxx if spans

 

 

 

are equal.

 

 

 

 

 

Upper limit

Set maximum and minimum

Upper limit: 105 % of operator

3.4.4

Lower limit

pressure alarm limits.

entered full scale pressure

 

 

 

Lower limit: None in absolute

 

 

 

and gauge modes. Minus

 

 

 

atmosphere in negative gauge

 

 

 

mode, minus equivalent of upper

 

 

 

limit in negative gauge mode on

 

 

 

a BG15K Q-RPT. Minus 66 kPa

 

 

 

(9.6 psi) with BA100K.

 

 

 

 

 

Resolution

Set display resolution

10 ppm of AutoRange span or

3.4.2

 

 

1 ppm of default Q-RPT span,

 

 

 

whichever is larger.

 

 

 

 

 

Stability Limit

Set Ready/Not Ready stability

50 ppm of AutoRange span or 2

3.4.3,

 

criterion.

ppm of default Q-RPT span,

3.3.4

 

 

whichever is larger.

 

 

 

 

 

Q-RPT turn down

Turn down Q-RPT

See RPM4 pressure

1.2.2.1

(Premium class Q-RPTs only)

characterization to improve

measurement specifications

 

 

measurement uncertainty when

 

 

 

span is reduced

 

 

 

 

 

 

Page 31

© 2004-2007 DH Instruments, a Fluke Company

Page 41