RPM4™ OPERATION AND MAINTENANCE MANUAL

READRATEn

Purpose

Read or set the specified or active Q-RPT read rate or auto read rate mode.

Command

“READRATEn period

 

Query

“READRATEn?”

 

 

Classic

“READRATEn= period

 

 

“READRATEn

 

 

Optional Suffix

“n”

 

The active Q-RPT is assumed if no suffix is given.

 

 

 

‘1’

Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF and the HL

 

 

 

 

Q-RPT is not active.

 

 

 

 

Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF

 

 

 

‘2’

Specify the Lo Q-RPT if the HL Q-RPT is not active and measurement

 

 

 

 

mode is not differential.

 

 

 

‘3’

Specify the HL Q-RPT if the HL Q-RPT is active.

Arguments

“period ":

Period to integrate the measurement over (ms). Can be from 200 ms to 20000

 

 

 

ms. Set to ‘0’ to enable automatic read rate.

Remarks

The RPM4 can have the speed in which it integrates each measurement (“read rate”) automatically

 

adjusted based on the rate, or it can manually set to a fixed period. Increasing the read rate increases

 

the reply time for pressure and rate queries. In differential mode and when using the HL Q-RPT, the Lo

 

Q-RPT read rate is always set to the same read rate as the Hi.

Example

Cmd sent:

“READRATE 1000”

(enhanced)

Query reply:

“1000”

(No reply if GPIB-488)

Example

Cmd sent:

“READRATE? 1000”

(enhanced)

Query reply:

“1000”

 

Example

Sent:

 

“READRATE=1000”

(classic)

Reply:

 

“1000”

 

Errors

ERR# 6:

The “period” argument is out of range.

 

ERR#10:

The suffix ‘n’ is invalid

See Also

“PR”, “PRR”, “SR”

 

 

 

3.5.7.2

 

 

 

 

 

 

 

 

 

 

READYCKn

 

 

 

 

 

Purpose

Read or set the Ready check flag.

Command

“READYCKn 1”

 

 

Query

“READYCKn?”

 

 

Classic

“READYCKn=1”

 

 

 

“READYCKn

 

 

Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given.

 

 

‘1’

Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF

 

 

 

Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF

 

 

‘2’

Specify the Lo Q-RPT

 

 

‘3’

Specify the HL Q-RPT if the HL Q-RPT is active

Remarks

The internal ready check flag is cleared whenever the specified or active Q-RPT reaches a

 

Not Ready (NR) condition. The “READYCK” query returns the status of the flag. The flag is

 

set by sending the “READYCK 1” program message while the Q-RPT is in a Ready condition.

 

The “READYCK” program message query can then be used at a later time to determine

 

whether a Not Ready condition has occurred since the ready check flag was set.

Example

Cmd sent:

“READYCK1 1”

(enhanced)

Query reply:

“1”

(no reply if GPIB-488)

Example

Cmd sent:

“READYCK1?”

(enhanced)

Query reply:

“1”

 

 

Example

Cmd sent:

“READYCK?”

(enhanced)

Query reply:

“1”

(if

Q-RPT condition has stayed Ready)

 

 

 

“0” (if Q-RPT condition has NOT stayed Ready)

Example

Sent:

 

“READYCK=1”

(classic)

Query reply:

“READYCK=1”

Example

Sent:

 

“READYCK”

(classic)

Query reply:

“READYCK=1” (if Q-RPT condition has stayed Ready)

 

 

 

“READYCK=0” (if Q-RPT condition has NOT stayed Ready)

Errors

ERR# 6:

Argument is not a ‘0’ or a ‘1’.

See Also

“SR”

 

 

 

 

 

3.2.1

 

 

 

 

© 2004-2007 DH Instruments, a Fluke Company