1. INTRODUCTION

Q-RPTs of A10M and lower are available with two different performance levels, STANDARD class and PREMIUM class. See the product label on the RPM4 rear panel, the Q-RPT module label on the rear panel and/or the product calibration reports to determine the class of the Q-RPT(s) installed in RPM4.

Table 1. Reference Pressure Transducer (Q-RPT) Module Designations and Ranges

 

 

 

 

SI VERSION

 

Q-RPT

 

 

MAXIMUM

 

 

 

MAXIMUM

 

 

DESIGNATION

 

 

RANGE

 

 

 

RANGE

 

 

 

 

 

[kPa]

 

 

 

[kPa]

 

 

 

 

 

Absolute

 

 

 

Gauge

 

 

 

 

 

 

 

 

 

 

 

 

A280M-L1

280 000

 

280 000

 

 

A200M-L1

200 000

 

200 000

 

 

A140M-L1

140 000

 

140 000

 

 

A100M-L1

100 000

 

100 000

 

 

A70M1

70 000

 

70 000

 

 

A40M1

40 000

 

40 000

 

 

A20M1

20 000

 

20 000

 

 

A14M1

14 000

 

14 000

 

 

A10M1

10 000

 

10 000

 

 

A7M1

7 000

 

7 000

 

 

A3.5M1

3 500

 

3 500

 

 

A2M1

2 000

 

2 000

 

 

A1.4M1

1 400

 

1 400

 

 

A700K1

700

 

700

 

 

A350K1

350

 

250

 

 

A200K1

200

 

100

 

 

A160K1

160

 

60

 

 

A100K1

110

 

10

 

 

BA100K4

110

 

--

 

 

G200K2

--

 

200

 

 

G100K2

--

 

100

 

 

G15K2

--

 

15

 

 

BG15K3

--

 

±15

 

 

 

 

 

 

 

 

 

 

 

US VERSION

MAXIMUM

MAXIMUM

RANGE

RANGE

[psi]

[psi]

Absolute

Gauge

 

 

40 000

40 000

30 000

30 000

20 000

20 000

15 000

15 000

10 000

10 000

6 000

6 000

3 000

3 000

2 000

2 000

1 500

1 500

1 000

1 000

500

500

300

300

200

200

100

100

50

35

30

15

23

8

16

1.5

16--

--30

--15

--2.2

--± 2.2

1.All AXXXX RPTs support absolute, gauge and compound (negative) gauge modes.

2.All GXXXX RPTs are positive gauge mode only.

3.BG15K is bi-directional gauge from - 15 to + 15 kPa (- 2.2 to + 2.2 psi).

4.BA100K is a barometric range whose low point is 70 kPa absolute (10 psi).

Page 3

© 2004-2007 DH Instruments, a Fluke Company

Page 13
Image 13
Fluke RPM4 manual SI Version, Rpt, Designation, US Version, 000 500 300 200 100