4. REMOTE OPERATION

REMOTE

Purpose

Lock out the front panel keypads during remote operation.

 

Command

“REMOTE”

 

 

 

Classic

“REMOTE”

 

 

 

Remarks

The RPM4 goes into remote mode whenever communications take place.

The user can

 

return to local operation by pressing the [ESC] key. The REMOTE program message locks

 

out the front panel completely. The only way to unlock the front panel after the “REMOTE”

 

command is using the “LOCAL” program message, the IEEE-488 “GTL” command, or by

 

cycling the RPM4 power.

 

 

Example

Cmd sent:

“REMOTE”

 

 

(enhanced)

Reply:

 

“REMOTE” (no reply if IEEE-488)

 

Example

Cmd sent:

“REMOTE?”

 

 

(enhanced)

Reply:

 

“REMOTE”

 

 

Example

Sent:

 

“REMOTE”

 

 

(classic)

Reply:

 

“REMOTE”

 

 

See Also

“LOCAL”

 

 

 

 

 

 

 

 

 

RESn

 

 

 

 

 

Purpose

To read or set the pressure display resolution for the active range.

 

Command

“RESn res

 

 

 

Query

“RESn

 

 

 

 

Classic

“RESn= res

 

 

 

“RESn

 

 

 

 

Default

“RESn=0.001”

 

 

Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given.

 

 

 

‘1’

Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF

 

 

 

 

Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF

 

 

 

‘2’

Specify the Lo Q-RPT

 

 

 

 

‘3’

Specify the HL Q-RPT if the HL Q-RPT is active

 

 

 

 

 

Arguments

res:

 

The pressure display resolution in % span of the specified Q-RPT range (0.0001 to 1 %

 

 

 

FS).

 

 

Remarks

The pressure display resolution is defined as % span of the active range.

The setting is

 

separate for each range, and changes as the range is changed.

 

Example

Cmd sent:

“RES .01”

 

 

(enhanced)

Query reply:

“0.01” (No reply if GPIB-488)

 

Example

Cmd sent:

“RES? .01”

 

 

(enhanced)

Query reply:

“0.01”

 

 

Example

Sent:

 

“RES=.01”

 

 

(classic)

Reply:

 

“0.01”

 

 

Errors

ERR# 6

The argument is invalid.

 

See Also

3.4.2

 

 

 

 

 

 

 

 

 

 

RESET

 

 

 

 

 

Purpose

Reset the user’s settings to factory defaults.

 

Command

“RESET”

 

 

 

Classic

“RESET”

 

 

 

Remarks

The RPM4 has user settings (units, resolution, etc.) that can be reset to factory defaults. The

 

remote “RESET” program message corresponds to the front panel “Reset - Sets”. System

 

calibration coefficients and communications settings are not affected. The reset cycle takes up

 

to 3 seconds to complete. Remote communications should not take place during this period.

Example

Cmd sent:

“RESET”

 

 

(enhanced)

Reply:

 

“RESET”

(no reply if IEEE-488)

 

Example

Cmd sent:

“RESET?”

 

 

(enhanced)

Reply:

 

“RESET”

 

 

Example

Sent:

 

“RESET”

 

 

(classic)

Reply:

 

“RESET”

 

 

See Also

3.5.9.1

 

 

 

 

Page 101