7.1REMOTE [ENT]

The RPM4 remote ENTER function operates by detecting the open or closed status of the REMOTE ENTER switch.

The normal status of the ENTER switch is OPEN. When RPM4 detects a CLOSED condition held for 300 ms, it is interpreted as equivalent to a press of the [ENT] key.

Any switch may be used to accomplish the remote ENTER function. The switch should be installed on the optional remote ENTER cable (DHI P/N 103128). Install the switch by connecting its two terminals to the black and white wires of the remote ENTER cable.

Figure 10. Remote [ENT] Connector Schematic

1. Black wire

2. White wire