3. OPERATION

3.2.8 DIRECT FUNCTION KEYS SUMMARY

Table 5 provides a brief summary of direct function key operation. It may be useful to keep a copy of this summary near the RPM4, especially when first becoming acquainted with its operation.

Local operation of RPM4 is through the front panel keypad. To minimize multi-layered menu structures, the keypad numerical keys also provide direct access to the most commonly used functions. The function accessed is labeled on the bottom half of the key. Direct function keys are active whenever RPM4 is in its MAIN RUN screen. Table 5 summarizes the operation of the direct function keys.

Table 5. Summary of RPM4 Function Key Operation

DIRECT FUNCTION KEYS ARE ACTIVE FROM THE MAIN RUN SCREEN

See corresponding manual sections for full detail.

 

Adjust height of fluid head

 

View active range and toggle

 

correction. Set to zero to defeat

 

through available ranges. [ENT] on

 

correction.

 

 

 

a range activates it.

 

 

 

 

 

 

 

.SDS controls the connection of

 

Select pressure unit of measure for

 

the Q-RPT to the TEST(+) port.

 

the active

range. Choice

of units

 

Close any open SDS. Open SDS

 

available in this menu can be

 

of active Q-RPT if closed.

 

 

customized.

 

 

 

 

 

 

 

 

 

 

Run the AutoZ function to

 

Select

the

pressure

measurement

 

AutoZero the active Q-RPT and

 

mode for the active range (absolute,

 

measurement mode.

 

 

gauge, negative gauge, differential).

 

 

 

 

 

 

 

 

 

 

 

 

 

Set

up

and

optimize

RPM4

 

Menu of less commonly used internal

 

features

for a

specific operating

 

 

 

functions and settings.

 

 

range and measurement mode.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Menu of commonly used setup

 

Run

automated leak

checking

 

features

 

including

save/delete

 

 

AutoRange

ranges,

set

display

 

routine.

 

 

 

 

 

 

 

 

 

resolution, set stability limit, set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

upper/lower, view AutoTest data

 

 

 

 

 

 

 

 

 

Select the display function for the

 

Start or set functions such as Leak

 

 

Check,

AutoTest

and

Freeze.

 

bottom line of the RPM4 display.

 

 

 

ENTER values when editing.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.3DIRECT FUNCTION KEYS

3.3.1 [RANGE]

PURPOSE

To view and/or change the active pressure measurement range and associated settings.

PRINCIPLE

An RPM4 range is defined as a pressure span and associated settings (see Section 3.2.3).

An RPM4 may have multiple ranges. Each Q-RPT in the RPM4 system has a default range which is its maximum span. Additional ranges, lower than the Q-RPT’s maximum span, may also be created using AutoRange (see Section 3.3.4). Ranges created using AutoRange may be saved, with their settings, for reactivation (see Section 3.4.1).

The [RANGE] function key allows the available RPM4 pressure ranges to be viewed and selected.