RPM4™ OPERATION AND MAINTENANCE MANUAL

UDU

Purpose

Read or set the user defined pressure unit.

Command

“UDU label, ucoef

 

 

Query

“UDU?”

 

 

 

Classic

“UDU=label, ucoef

 

 

 

“UDU”

User unit label (4 alphanumeric char maximum). It cannot be an already

 

label:

 

 

supported unit label.

 

 

Ucoef:

“User unit conversion coefficient (units/Pa).

Default

“UDU USER,1.0”

 

 

Remarks

The user defined unit must be set up with the program message prior to remote or local

 

selection.

 

 

 

Example

Cmd Sent:

“UDU MYUN, .001”

 

(enhanced)

Query reply:

“MYUN, 0.001000”

(No reply if GPIB-488)

Example

Cmd Sent:

“UDU? MYUN, .001”

 

(enhanced)

Query reply:

“MYUN, 0.001000”

 

Example

Sent:

“UDU=MYUN, .001”

 

(enhanced)

Reply:

“MYUN, 0.001000”

 

Errors

 

 

 

 

See Also

3.5.6, 3.3.2

 

 

 

 

 

 

 

 

ULn

 

 

 

 

Purpose

Read or set an upper limit for the specified or active Q-RPT.

Command

“Uln limit

 

 

 

Query

“Uln?”

 

 

 

Classic

“Uln=limit

 

 

 

 

“ULn

 

 

 

Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given.

 

 

‘1’

Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF and the HL

 

 

 

Q-RPT is not active.

 

 

 

 

Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF

 

 

‘2’

Specify the Lo Q-RPT if the HL Q-RPT is not active and measurement

 

 

 

mode is not differential.

 

 

‘3’

Specify the HL Q-RPT if the HL Q-RPT is active.

Arguments

“limit":

The upper limit pressure in the specified Q-RPT’s current pressure unit and

 

 

measurement mode. If the unit is meters, this value is be interpreted in kPa. If

 

 

the unit is feet, this value is be interpreted in psi.

Remarks

The RPM4 has an upper limit for each range and for each measurement mode (gauge, absolute, and

 

differential).

If the pressure does exceed the upper limit, the pressure display flashes. This

 

feature should always be used to prevent accidental over pressure of a device under test.

Example

Cmd sent:

“UL 1000”

 

(enhanced)

Query reply:

“1000.00 kPa a” (No reply if GPIB-488)

Example

Cmd sent:

“UL? 1000”

 

(enhanced)

Query reply:

“1000.00 kPa a”

 

Example

Sent:

“UL=1000”

 

(classic)

Reply:

“1000.00 kPa a”

 

Errors

ERR# 6:

The “limit” argument is out of range.

See Also

“LL”

 

 

 

 

3.4.4

 

 

 

© 2004-2007 DH Instruments, a Fluke Company

Page 106

Page 116
Image 116
Fluke RPM4 manual ULn