3. OPERATION

 

 

 

 

 

 

Table 11. Reset – Sets

 

 

 

 

 

 

RESET

RESULT

SEE SECTION

 

[UNIT]

Pressure unit of measure to first of the six available

3.3.2

 

 

 

 

 

[MODE]

Measurement mode to the native mode of the Q-RPT

3.3.3

 

 

 

 

 

[RANGE]

Hi Q-RPT default range

3.3.1

 

 

 

 

 

[HEAD]

0 cm height and Nitrogen medium

3.3

 

 

 

 

 

Stability Limit

0.005 % of Hi Q-RPT default range

3.4.3

 

 

 

 

 

Upper Limit

Hi Q-RPT default range default value

3.4.4

 

 

 

 

 

Resolution

0.001 % FS of Hi Q-RPT default range

3.4.2

 

 

 

 

 

AutoZ

AutoZ ON for all Q-RPTs and measurement modes.

3.5.1

 

 

Zoffset value not affected

 

 

AutoZ

Patm,0 set to 101.325 kPa a

3.5.1, 3.2.2

 

 

 

 

 

Leak Check

15 second run time. Clear logged results

3.3.5

 

 

 

 

 

Screen Saver

10 minutes to activation

3.5.5.1

 

 

 

 

 

Key Sounds

Medium tone valid key press sound

3.5.5.2

 

 

 

 

 

Lo Vnt

Automatic

3.5.7.4

 

 

 

 

 

ReadRate

Automatic

3.5.7.2

 

 

 

 

3.5.9.2<2 UNITS>

PURPOSE

Sets the six pressure units available under the UNIT function to the SI or US default selections depending on whether the RPM4 has been factory set for SI or US (see Section 3.3.2).

Sets the user defined unit to 1.000/Pa (see Section 3.5.6). Sets the reference temperature for inWa unit to 20°C.

3.5.9.3<3ATEST>

PURPOSE

Resets Auto Tests to default characteristics (see Section 3.3.10).

Clears AutoTest data log (see Section 3.4.5).

3.5.9.4<4 CAL>

PURPOSE

The Reset - Cal function will reset Q-RPT, utility sensor and barometer calibration coefficients and settings and reset AutoZ values to zero. This will change the RPM4 calibration and could cause it to make out of tolerance measurements.

Clears all user values affecting the calibration of Q-RPTs and the on-board barometer. The Reset – Cal resets are itemized in Table 12.

Page 79

© 2004-2007 DH Instruments, a Fluke Company

Page 89
Image 89
Fluke RPM4 manual 9.2 2 Units, 9.3 3ATEST, 9.4 4 CAL, Reset Result See Section Unit