RPM4™ OPERATION AND MAINTENANCE MANUAL

PRESSURE1UNITM hzRR

DDISPLAYFUNCIONnn/nn

DISPLAY FIELD

NAME

PURPOSE

 

CONTENTS

SECTION

1.

PRESSURE1

Measured

Displays pressure

Numerical pressure value and sign.

1.2.2.1

 

 

pressure

measured by the

Intermittently flashes <SDS

3.2.7

 

 

 

active Q-RPT

CLOSED> when the Q-RPT is shut

 

 

 

 

 

 

 

 

off from the TEST(+) port by SDS.

 

2.

UNIT

Unit of

Identifies unit of

Pressure unit abbreviation

3.3.2

 

 

measure

measure in which

 

 

 

 

 

 

pressure values are

 

 

 

 

 

 

displayed

 

 

 

3.

M

Measurement

Identifies

<a>

absolute

3.3.3

 

 

mode

measurement mode

<g>

gauge or negative gauge

 

 

 

 

of displayed pressure

<d>

differential

 

 

 

 

 

 

 

 

4.

h

Head

Indicates whether a

<h>

the fluid head is not zero

3.3.7

 

 

pressure

fluid head correction

<blank> fluid head is zero

 

 

 

indicator

is applied to

 

 

 

 

 

 

PRESSURE1

 

 

 

5.

z

AutoZero

Indicates whether

<z>

AutoZ is ON

3.5.1

 

 

indicator

the AutoZero function

<blank> AutoZ is OFF

 

 

 

 

is ON or OFF for the

 

 

 

 

 

 

active Q-RPT and

 

 

 

 

 

 

measurement mode

 

 

 

6.

RR

Active

Indicates the position

<Hi>

Hi Q-RPT

3.2.3

 

 

Q-RPT

of the active Q-RPT

<Lo> Lo Q-RPT

 

 

 

position

in the RPM4

<HL> Hi and Lo Q-RPTs together in

 

 

 

indicator

 

 

parallel measurement mode

 

 

 

 

 

<Hd> Hi and Lo Q-RPTs together in

 

 

 

 

 

 

differential measurement

 

 

 

 

 

 

mode

 

 

 

 

 

 

 

7.

nn/nn

Sequence

Indicates progress of

<NN/NN> Number of this point

3.3.10

 

 

progress

ATest sequence,

 

over total number of points

 

 

 

indicator

during test execution

 

in the sequence

 

8.

DISPLAY

Information

Pressure indication

Numerical pressure value

3.3.6

 

FUNCTION

specific to

depending on current

and sign.

 

 

 

the DISPLAY

RPM4 DISPLAY

<PPC3 EXT DEV:> when the RPM4

 

 

 

mode

function. Leading

is initialized as an external device

 

 

 

 

character identifies

to a PPC3 pressure controller.

 

 

 

 

the value

 

 

 

9.

D

Pressure

Pressure information

<σ>

Display mode is AVERAGE

3.3.6

 

 

information

indicator depending

 

and value is standard deviation

 

 

 

indicator

on current RPM4

<R>

Display mode is RATE and

 

 

 

 

DISPLAY function.

 

value is pressure rate of

 

 

 

 

 

 

change per second

 

 

 

 

 

<H>

Display mode is HI/LO and

 

 

 

 

 

 

value is high, then low

 

 

 

 

 

<D>

Display mode is DEVIATION

 

 

 

 

 

 

and value is difference from

 

 

 

 

 

 

current target

 

 

 

 

 

<Σ, , > Display mode is RPT

 

 

 

 

 

 

and value is measurement

 

 

 

 

 

 

of inactive RPT

 

 

 

 

 

<F>

Display mode is FREEZE and

 

 

 

 

 

 

value is last captured reading

 

 

 

 

 

Blank, no character Current

 

 

 

 

 

 

display mode is CLEAN

 

 

 

 

 

 

 

 

Figure 3. MAIN RUN Screen Display Fields

© 2004-2007 DH Instruments, a Fluke Company

Page 16

Page 26
Image 26