6. TROUBLESHOOTING

 

 

 

 

 

 

 

 

 

 

 

SYMPTOM

PROBABLE CAUSE

SOLUTION

 

 

Disagreement between two Q-RPTs

Difference is actually within tolerance and

Compare differences observed to

 

 

tolerances on Q-RPT measurements.

 

 

in system appears excessive.

acceptable disagreement.

 

 

1.2.2.1

 

 

 

 

 

 

Apparent inaccurate pressure

Reference transducer destroyed by

Contact DHI Authorized Service Provider.

 

 

control/measure and little or no

 

 

overpressure.

Table 29

 

 

response from Q-RPT:

 

 

 

 

 

 

 

 

 

 

 

 

Incorrect pressure units and/or measurement

Set desired pressure units and/or

 

 

 

measurement mode. Consider reference

 

 

 

mode (gauge or absolute).

 

 

 

temperature if unit is inWa. 3.3.2, 3.3.3

 

 

Apparent inaccurate pressure

 

 

 

 

 

 

 

Q-RPT calibration coefficients have been

Check and correct calibration coefficients if

 

 

measurement.

 

 

altered.

needed. 5.2

 

 

 

 

 

 

 

 

 

 

 

AutoZ has been run and turned ON with an

Check value of Poffset. Rerun AutoZ with a

 

 

 

incorrect standard for zero.

valid reference. 3.5.1

 

 

Apparent inaccurate pressure

An unplanned “head” correction is active or

Operation is normal. Remove or change

 

 

measurement and <h> is

 

 

head height or fluid is incorrect.

“head” correction. 3.3.7

 

 

displayed on top line of screen.

 

 

 

 

 

 

 

Need to run AutoZ to rezero in gauge mode.

Run AutoZ. 3.3.9

 

 

Pressure applied is zero gauge

 

 

 

 

Current measurement mode is absolute and

Check measurement mode setting and

 

 

but reading is NOT zero.

set to gauge if gauge pressure

 

 

 

RPM4 is indicating atmospheric pressure.

measurements are desired. 3.3.3

 

 

 

 

 

 

 

 

 

 

 

 

 

Operation is normal. Check

 

 

Pressure applied is atmospheric

Current measurement mode is gauge and

measurement mode setting and set to

 

 

but RPM4 indicates near zero.

RPM4 is indicating zero gauge pressure.

absolute if absolute pressure

 

 

 

 

measurements are desired. 3.3.3

 

 

A third Q-RPT designated HL is

Parallel measurement mode has been

Operation is normal. Turn off parallel

 

 

available but RPM4 has only two

 

 

activated.

measurement mode if desired. 3.5.7.3

 

 

Q-RPTs.

 

 

 

 

 

 

 

 

 

 

 

NOT reading pressure applied to

SDS is CLOSED for the active Q-RPT so the

After taking precautions to assure that

 

 

the test port and <SDS CLOSED>

 

 

pressure applied to the test port is NOT

pressure applied to test port is safe for

 

 

is flashing over the pressure

 

 

getting to the Q-RPT.

the RPT, OPEN SDS. 3.2.7, 3.3.8

 

 

indication.

 

 

 

 

 

 

 

 

 

 

 

 

RPM4 may have two TEST(+) ports and the

Familiarize yourself with your RPM4

 

 

 

configuration. Connect pressure to be

 

 

 

pressure to be measured is connected to the

 

 

 

measured to correct TEST(+) port.

 

 

 

wrong Q-RPT’s port

 

 

Is NOT reading pressure applied

2.3.5, 5.6

 

 

 

 

 

to the TEST(+) port.

RPM4 may have two TEST(+) ports and

Familiarize yourself with your RPM4

 

 

 

pressure to be measured is connected to

configuration. Select active range to be

 

 

 

correct port but active Q-RPT is NOT the

on the Q-RPT to which the pressure to

 

 

 

Q-RPT on that TEST port.

be measured is corrected. 5.6, 3.3.1

 

 

<SDS CLOSED> is displayed

SDS is CLOSED for the active Q-RPT.

Operation is normal. OPEN SDS if

 

 

alternating with the measured

 

 

desired. 3.3.8

 

 

pressure.

 

 

 

 

 

 

 

SDS should be present on a

SDS has been turned full time OFF for the

 

 

 

Q-RPT but does not seem to be

Turn SDS full time ON. 3.5.4.2

 

 

Q-RPT.

 

 

active.

 

 

 

 

 

 

 

 

 

 

 

 

SDS will not OPEN for the Q-RPT

When not in differential or parallel

Use direct SDS direct control to OPEN

 

 

measurement mode, SDS on the inactive

 

 

shown on the second line in the

 

 

Q-RPT can only be OPENED using direct

SDS on the inactive Q-RPT. 3.5.4.1

 

 

Q-RPT DISPLAY function.

 

 

SDS control.

 

 

 

 

 

 

 

 

 

 

 

 

SDS was left OPEN on one

When changing ranges causes a change of

 

 

 

Q-RPT when ranges were

 

 

 

the active Q-RPT, SDS is always CLOSED

Operation is normal. 3.2.7

 

 

changed. When you come back

 

 

for the Q-RPT that becomes inactive.

 

 

 

to it, its CLOSED again.

 

 

 

 

 

 

 

 

 

 

 

 

You would like to eliminate the

Set SDS to be inactive except for an

Turn SDS full time OFF for the Q-RPT.

 

 

SDS function for a Q-RPT.

overpressure situation.

3.5.4.2

 