Fluke RPM4 manual SDSAUTOn

Models: RPM4

1 150
Download 150 pages 5.32 Kb
Page 113
Image 113

 

 

 

 

 

4. REMOTE OPERATION

 

 

 

 

 

 

 

 

 

 

 

 

 

SDSACT

 

 

 

 

 

Purpose

Read or set the status for both the Hi and Lo Q-RPT’s SDS systems. This is equivalent to

 

 

use of the [SDS] key from the RPM4 front panel.

 

Command

“SDS state

 

 

 

Query

“SDS”

 

 

 

 

Classic

“SDS=state

 

 

 

 

“SDS”

 

 

 

 

Default

“SDS=1”

 

 

 

Arguments

”state”:

 

‘0’ OPEN the SDS system on the active Q-RPT and CLOSE the SDS

 

 

 

 

system on the “inactive” Q-RPT.

 

 

 

 

‘1’ to CLOSE the SDS system on both Q-RPTs.

 

Remarks

This command can be used to select one Q-RPT at a time, ensuring that the “inactive” Q-RPT

 

 

SDS is closed. If a Q-RPT’s SDS is full time “OFF”, then it is not be affected by this

 

 

command. The status of the active Q-RPT’s SDS system is replied. If the RPM4 is in

 

 

differential mode or the HL Q-RPT is active, then this command opens or close both the Hi

 

 

and Lo SDS at the same time, unless the SDS is full time “off”.

 

Example

Cmd sent:

“SDS? 1”

 

 

(enhanced)

Query reply:

“1”

 

 

Example

Cmd sent:

“SDS=0”

 

 

(classic)

Query reply:

“SDS=0”

 

 

Errors

ERR# 7:

Argument not a ‘0’ or a ‘1’

 

 

ERR# 10:

Invalid suffix or command argument.

 

 

ERR# 23:

SDS not installed on this Q-RPT

 

See Also

“SDSAUTO”, “SDS”

 

 

 

3.2.7, 3.3.8, 3.5.4

 

 

 

 

 

 

 

 

SDSAUTOn

 

 

 

 

 

Purpose

Read or set the automatic full time ON/OFF status for the specified or active Q-RPT SDS

 

 

systems. This is equivalent to use of the SDS full time ON/OFF function from the RPM4

 

 

front panel.

 

 

 

Command

“SDSAUTOn on

 

 

Query

“SDSAUTOn

 

 

Classic

“SDSAUTOn=on

 

 

 

“SDSAUTOn

 

 

Default

“SDSAUTOn=1”

 

 

Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given. If the “HL” Q-RPT is the active Q-

 

 

 

RPT, and no suffix is specified, then the “Hi” Q-RPT is assumed

 

 

 

‘1’

Specify the Hi Q-RPT SDS

 

 

 

‘2’

Specify the Lo Q-RPT SDS

 

Arguments

on:

 

‘0’ to turn the SDS system OFF.

 

 

 

 

‘1’ to turn the SDS system ON

 

Remarks

Turning an SDS system to full time “OFF” or “ON” affects how the remote command

 

 

“SDSACT” and the front panel SDS button can affect each Q-RPT’s SDS system.

 

Example

Cmd sent:

“SDSAUTO2? 1”

 

 

(enhanced)

Query reply:

“1”

 

 

Example

Cmd sent:

“SDSAUTO1=0”

 

 

(classic)

Query reply:

“SDSAUTO1=0”

 

 

Errors

ERR# 7:

Argument not a ‘0’ or a ‘1’

 

 

ERR# 10:

Invalid suffix.

 

 

 

ERR# 23:

SDS not installed on this Q-RPT

 

See Also

“SDSACT”, “SDS”

 

 

 

3.2.7, 3.3.8, 3.5.4,

 

 

 

 

 

 

 

 

SCRSAV

 

 

 

 

 

Purpose

Read or set the front panel display screen saver activation time.

 

Command

“SCRSAV n

 

 

Query

“SCRSAV?”

 

 

 

Classic

“SCRSAV=n

 

 

 

“SCRSAV”

 

 

 

Arguments

n:

 

The inactivity period (0 – 99 minutes) after which screen saver activates.

 

Default

“SCRSAV 10”

 

 

Remarks

The RPM4 front panel will dim after a period of keyboard and remote inactivity. Setting this

 

 

value to ‘0’ disables this feature.

 

 

Example

Cmd sent:

“SCRSAV 30”

 

 

(enhanced)

Query reply:

“30”

(No reply if GPIB-488)

 

Example

Cmd sent:

“SCRSAV? 30”

 

 

(enhanced)

Query reply:

“30”

 

 

Example

Sent:

 

“SCRSAV=30”

 

 

(classic)

Reply:

 

“30”

 

 

Errors

ERR# 6

 

The argument was invalid.

 

See Also

3.5.5.1

 

 

 

Page 103

© 2004-2007 DH Instruments, a Fluke Company

Page 113
Image 113
Fluke RPM4 manual SDSAUTOn