RPM4™ OPERATION AND MAINTENANCE MANUAL

AutoZ in gauge measurement mode with a Gxxx or BGxxx Q-RPT

Q-RPTs with the designation Gxxx or BGxxx are intrinsically gauge. They have an atmospheric (TEST(-)) reference port and measure gauge pressure (difference from atmosphere).

In gauge measurement mode the value of Pstd,0 is always zero gauge pressure. Zero gauge pressure, by definition, is applied to the Q-RPT when it is vented to atmosphere.

When using a Gxxx Q-RPT in gauge measurement mode, with AutoZ ON, gauge pressure is calculated as:

Pgauge = Pu - Poffset,G

In gauge measurement modes, AutoZ is run when initiated by the operator pressing [AutoZ] when the RPM4 is vented. This assures the regular automated updating of the Poffset,G value.

AutoZ in differential measurement mode

RPM4s with two Q-RPTs of the same type (Axxx, Gxxx, BGxxx) support differential measurement mode (see Section 3.3.3.1) in which the measured pressure is the difference between the pressure measured by the Hi Q-RPT and the Lo Q-RPT (Hi – Lo).

In differential measurement mode, the purpose of AutoZ is to compensate for the offset between the Hi and Lo Q-RPT indications at the “line pressure” at which the differential pressure is being measured. The value of Pstd,0 is the pressure indicated by the Lo Q-RPT at the time of AutoZ.

When using an RPM4 with two Q-RPTs in differential measurement mode, with AutoZ ON, differential pressure is calculated as:

Pdifferential = PHi - PLo - Poffset,D

In differential measurement mode, AutoZ is run when initiated by the operator pressing [AutoZ] when both Q-RPTs of the RPM4 are connected to a common pressure..

AutoZ ON/OFF

The AutoZ function can be turned ON and OFF, separately for gauge, absolute and differential measurement modes. Table 7 summarizes the effect of AutoZ ON and OFF.

Table 7. AutoZ ON and OFF

 

Q-RPT

 

 

MEASUREMENT

 

 

AutoZ

 

 

Poffset

 

 

ΔPatm

 

AutoZ

 

 

 

 

 

 

 

 

 

 

ROUTINE

 

TYPE

 

 

MODE

 

 

STATUS

 

 

APPLIED

 

 

APPLIED

 

 

 

 

 

 

 

 

 

 

 

RUNS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ON

 

YES

 

 

 

When initiated

 

 

 

 

Absolute

 

 

 

Not Applicable

by operator

 

 

 

 

 

 

 

 

 

 

 

 

Axxx

 

 

 

 

 

OFF

 

NO

 

 

 

Not available

 

 

 

 

Gauge or

 

ON

 

YES

 

YES

When initiated

 

 

 

 

negative gauge

 

 

 

 

 

 

 

by operator

 

 

 

 

 

OFF

 

 

NO

 

 

 

 

 

 

 

 

 

 

 

 

 

Gxxx

 

 

Gauge or

 

ON

 

YES

 

 

 

When initiated

 

or

 

 

 

 

 

 

 

 

 

 

 

by operator

 

BGxxx

 

 

negative gauge

 

 

 

 

 

 

 

Not Applicable

 

 

 

 

 

OFF

 

NO

 

Not available

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Any, Hi Q-RPT

 

 

 

 

 

ON

 

YES

 

 

 

When initiated

 

 

 

Differential

 

 

 

Not Applicable

by operator

 

in dual Q-RPT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RPM4

 

 

 

 

 

OFF

 

NO

 

 

 

Not available

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Recommendations for the Use of the AutoZ Function

The AutoZ function provides a powerful and easy to use tool for improving the stability over time of RPM4 Q-RPTs and maximizing the recalibration interval by compensating for change in zero between full recalibrations. In differential measurement mode, it minimizes the uncertainty in the differential pressure due to the offset between the Hi and Lo Q-RPTs at the