RPM4™ OPERATION AND MAINTENANCE MANUAL

Table 12. Reset – Cal

RESET

RESULT

SEE SECTION

All Q-RPT Calibration

PA to zero, PM to 1

5.2.1.1, 5.2.7

Coefficients

 

 

 

 

 

Axxx Q-RPT absolute and

ON

5.2.5

negative gauge mode

 

 

 

 

 

On-board Barometer

PA to zero, PM to 1

5.3

Calibration Coefficients

 

 

 

 

 

Calibration Date

Set all dates to 19800101

5.2.7, 5.3

 

 

 

AutoZ values

Axxx Q-RPTs: All Poffset values to zero for

3.5.1

 

absolute mode and 101325 Pa for gauge and

 

 

negative gauge modes

 

 

Gxxx, BGxxx Q-RPTs: All Poffset values to zero

 

 

 

 

AutoZ function

ON, all Q-RPTs, all measurement modes

3.5.1

 

 

 

3.5.9.5<5 ALL>

PURPOSE

The reset - all function clears and deletes large amounts of user defined information including critical calibration data.

Combines all resets in one global reset command that clears the entire user section of non-volatile memory except the ID function (see Section 3.5.5.4) and the security level password (see Section 3.5.5.5) returning RPM4 to the “as delivered” condition. The Reset – All resets are itemized in Table 13.

Table 13. Reset – All

RESET

RESULT

SEE SECTION

Reset – Sets

All the resets of Reset - Sets

3.5.9.1

 

 

 

Reset – Units

All the resets of Reset - Units

3.5.9.2

 

 

 

Reset – Atest

All the resets of Reset - ATest

3.5.9.3

 

 

 

Reset – Cal

All the resets of Reset - Cal

3.5.9.4

 

 

 

Remote Interfaces

COM1, COM2 and IEEE-R88 interfaces to

3.5.2

 

default settings.

 

 

 

 

Remote Communications

Remote command format to Classic

3.5.2.3

 

 

 

Level (Security)

Reset security level to low

3.5.5.5

 

 

 

Parallel Measurement Mode

Off

3.5.7.3

 

 

 

SDS

All SDS full time ON

3.5.4.2

 

 

 