RPM4™ OPERATION AND MAINTENANCE MANUAL

L2 / L3

Purpose

Read or set the type of program command format to use (classic or enhanced).

Command

“L2”enables “classic” mode

 

 

“L3”enables “enhanced” mode

 

Defaults

“L2”

 

(Classic mode)

 

Remarks

The user can select the type of remote command format using these simplified commands.

 

There is no query format. This format must agree with the format sent to the RPM4. This

 

command is a replacement for the “MSGFMT” command.

Example

Cmd sent:

“L3”

 

 

(enhanced)

Query reply:

“L3”

 

(No reply if IEEE-488)

Example

Sent:

 

“L2”

 

 

(classic)

Reply:

“L2”

 

 

See Also

“MSGFMT”

 

 

 

 

4.3, 3.5.2.3

 

 

 

 

 

 

 

 

 

LLn

 

 

 

 

 

Purpose

Read or set the lower pressure limit for the active range and measurement mode (negative

 

gauge or differential measurement mode only).

Command

“LLn

limit

 

 

 

Classic

“LLn =limit

 

 

 

Optional Suffix

“n”

 

The active Q-RPT is assumed if no suffix is given.

 

 

 

‘1’

Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF and the HL

 

 

 

 

Q-RPT is not active.

 

 

 

 

Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF

 

 

 

‘2’

Specify the Lo Q-RPT if the HL Q-RPT is not active and measurement

 

 

 

 

mode is not differential.

 

 

 

‘3’

Specify the HL Q-RPT if the HL Q-RPT is active.

Arguments

"limit:

The lower pressure limit for the current pressure range in the specified Q-

 

 

 

RPT’s current unit of measure unit and measurement mode. Value is always

 

 

 

a negative value of gauge pressure.

Remarks

Negative gauge and differential modes in RPM4 have a lower limit. If the pressure is less

 

than the lower limit, the pressure display flashes. This feature should always be used to

 

prevent accidental over (under) pressure of a device under test.

Example

Cmd sent:

“LL

-4”

 

(enhanced)

Reply:

“-4 kPa g”

(no reply if IEEE-488)

Example

Cmd sent:

“LL?

-4”

 

(enhanced)

Reply:

“-4 kPa g”

 

Example

Sent:

 

“LL=-4”

 

(classic)

Reply:

“-4 kPa g”

 

Errors

ERR# 6:

The limit argument is invalid.

 

ERR# 10

The suffix is invalid

 

ERR# 23

The mode must be negative gauge to specify a lower limit.

See Also

“UL”, “MMODE”

 

 

 

 

3.4.4,

3.3.3

 

 

 

 

 

 

 

 

 

LOCAL

 

 

 

 

 

Purpose

Returns control to the RPM4 front panel.

Command

“LOCAL”

 

 

 

Classic

“LOCAL”

 

 

 

Remark

The REMOTE program message can lock the front panel out completely. The user can return

 

to local operation by sending the LOCAL program message, sending the IEEE-488 ‘GTL’

 

command (if in enhanced format), or by cycling RPM4 power.

Example

Cmd sent:

“LOCAL”

 

(enhanced)

Reply:

“LOCAL”

(no reply if IEEE-488)

Example

Cmd sent:

“LOCAL?”

 

(enhanced)

Reply:

“LOCAL”

 

Example

Sent:

 

“LOCAL”

 

(classic)

Reply:

“LOCAL”

 

See Also

“REMOTE”

 

 

 

© 2004-2007 DH Instruments, a Fluke Company

Page 94

Page 104
Image 104
Fluke RPM4 manual L2 / L3, LLn =limit, Limit