Fluke RPM4 manual Calibration Point Sequence, Premium Class BGxxx Q-RPTs

Models: RPM4

1 150
Download 150 pages 5.32 Kb
Page 131
Image 131

5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS

Table 25. Calibration Point Sequence, Premium Class,

Axxx and Gxxx Q-RPTs

CALIBRATION

POINT

SEGMENT

NO.

 

1

 

2

 

3

 

4

Ascending

5

6

 

7

 

8

 

9

 

10

 

11

 

12

Descending

13

14

 

 

15

POINT

[% OF SPAN]

Lowest Pressure1

5%

10%

15%

20%

30%

40%

50%

60%

80%

100%

80%

50%

20%

Lowest Pressure1

1In absolute mode, lowest point that can be set reliably by the reference piston gauge. Zero (vented) in gauge mode.

Axxx (absolute) Q-RPTs that will NOT be used in absolute or negative gauge measurement modes may be calibrated in gauge measurement mode with a gauge pressure standard. Axxx (absolute) Q-RPTs calibrated in gauge measurement mode should be used in gauge mode only (see Section 5.2.5).

Table 26. Calibration Point Sequence, Premium Class,

BGxxx Q-RPTs

CALIBRATION

POINT

SEGMENT

NO.

 

1

 

2

 

3

 

4

Ascending

5

6

 

7

 

8

 

9

 

10

 

11

 

12

Descending

13

14

 

 

15

POINT

[% OF SPAN]1

0%

15%

30%

40%

45%

50%

55%

60%

70%

85%

100%

70%

50%

30%

0%

1.The BG15K span is – 15 to + 15 kPa, 30 kPa total. Therefore, 0 % is – 15 kPa, 25 % is – 7.5 kPa, 100 % is + 15 kPa, etc.

BGxxx (bi-directional gauge) Q-RPTs that will not be used in negative gauge measurement mode may be calibrated in gauge (positive pressures) measurement mode only. Use the Premium Class gauge mode (Gxxx) calibration point sequence (see Table 25). BGxxx Q-RPTs calibrated in gauge measurement mode only should not be used in negative gauge mode.

Page 121

© 2004-2007 DH Instruments, a Fluke Company

Page 131
Image 131
Fluke RPM4 Calibration Point Sequence, Premium Class BGxxx Q-RPTs, 15 % 30 % 40 % 45 % 50 % 55 % 60 % 70 % 85 % 100 %