RPM4™ OPERATION AND MAINTENANCE MANUAL

3.1.3 REMOTE [ENT] (ENTER) FOOTSWITCH

The optional remote ENTER function is a switch that duplicates the function of the front panel [ENT] key. The remote ENTER function is serviced by a connector on the RPM4 rear panel. An optional footswitch is available to activate remote entry hands free or a different switch may be used. See Section 7.1 for information on remote ENTER switch wiring.

The remote ENTER feature can be particularly convenient when running AutoTests (see Section 3.3.10) in which using a footswitch to [ENT] allows hands free operation. It can also be used, with the FREEZE display function (see Section 3.3.6.6) to capture the activation point of a pressure switch.

3.1.4 SOUNDS

RPM4 is equipped with a variable frequency tone device to provide audible feedback and alarms. The beeper is used for the following indications.

Valid key press

Brief beep. Choice between three frequencies or NO

 

sound is available (see Section 3.5.5.2).

Invalid key press

Descending two tone “blurp”

Leak check completed

Three two second beeps (see Section 3.3.5).

Upper or lower limit exceeded

Intermittent one second beeps (see Section 3.4.4).

Pmax! (overpressure limit)

Eight second high frequency beep (see Section 3.4.4.1).

exceeded

 

Parallel measurement mode

Rapid beeps for 8 seconds (3.2.4).

possible disconnection of the

 

Hi and Lo Q-RPTs

 

AutoTest in/out of tolerance

Ascending triad/descending triad (see Section 3.3.10).

reading

 

3.2GENERAL OPERATING PRINCIPLES

3.2.1 PRESSURE READY/NOT READY

There is a Ready/Not Ready indication LED on the RPM4 front panel. This indication is intended to provide the user with a clear and objective indication of when a stable pressure has been achieved. Ready is indicated when the current stability (rate of change) of pressure is less than the stability limit. The user can set the stability limit (see Section 3.4.3). The ready indication is often used when comparing the RPM4 and a test device to indicate when a valid reading can be made.

In RPM4s with two Q-RPTs, the Ready/Not Ready indication always applies to the active Q-RPT, whose pressure measurement is displayed on the top line of the RPM4 display. When the inactive Q-RPT measurement is displayed on the second line using the RPT display mode, the Ready/Not Ready indication for the inactive RPT is indicated by the leading character on the line. <Σ> indicates Ready. < > or < > indicate Not Ready, pressure increasing or pressure decreasing.

The Ready/Not Ready LED indications are:

<Green >

Pressure Ready

The pressure stability is within the stability limit.

<Red>

Pressure Not Ready

The pressure stability is NOT within the stability limit.