SECTION 11—BATTERIES

8.Connect joystick cable (not shown). Refer to Disconnecting/Connecting the Joystick on page 123.

Top Shroud

BLACK Battery

Battery Handle

Connectors

 

Rear Battery

RED Battery Connectors

 

 

BLACK Controller

Joystick

Connector

 

Cable

Plug into

 

Controller

Rear of

 

Wheelchair

Front

 

Wheelchair

Battery

 

Base Frame

 

Front of

Wheelchair

FIGURE 11.2 Removing/Installing Batteries from/into Battery Tray

Connecting/Disconnecting Battery Cables

Connecting Battery Cables

WARNING

NEVER allow any of your tools and/or battery cables to contact both battery terminals at the same time. An electrical short may occur and serious personal injury or damage may occur.

Connect same color connectors to each other (RED to RED, BLACK to BLACK).

DO NOT remove fuse or mounting hardware from POSITIVE (+) battery cable mounting screw. To replace the fuse, obtain and replace battery harness with fuse.

The POSITIVE (+) battery cable MUST connect to the POSITIVE (+) battery terminal, otherwise serious damage will occur to the electrical system.

The use of rubber gloves is recommended when working with batteries.

Part No. 1125075

101

Pronto® Series

Page 101
Image 101
Invacare M61, M50, M51 service manual Connecting/Disconnecting Battery Cables, Connecting Battery Cables