SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING

NUMBER OF FLASHES

DIAGNOSTICS CODE

ERROR CODE DESCRIPTION

SUB CODE*

DETAILS OF ERROR CODE

POSSIBLE SOLUTION

 

 

 

 

 

 

8

E08

Controller Fault

00

Controller fault

Check connections and wiring. Replace

 

 

 

 

 

power module.

 

 

 

01

RAM fault

 

 

 

 

 

 

 

 

 

 

 

 

 

02

ROM fault

 

 

 

 

 

 

 

 

 

 

03

CPU fault

 

 

 

 

 

 

 

 

 

 

04

EEPROM fault

 

 

 

 

 

 

 

 

 

 

05

Watchdog fault

 

 

 

 

 

 

 

 

 

 

06

Stack fault

 

 

 

 

 

 

 

 

 

 

07

Software fault

 

 

 

 

 

 

 

 

 

 

08

Power-up testing fault

 

 

 

 

 

 

 

 

 

 

09

Relay fault or precharge fault

 

 

 

 

 

 

 

 

 

 

10

Bridge fault or disable all fault

 

 

 

 

 

 

 

 

 

 

11

Electronics fault: Thermistor

 

 

 

 

 

 

 

 

 

 

12

Calibration setting fault

 

 

 

 

 

 

 

9

E09

Communications

00

Remote connection lost

Check connections and wiring. Replace

 

 

Fault

 

 

Bus cable.

 

 

01

Low communication mode

 

 

 

 

 

 

 

 

 

 

10

E10

General Fault

00

General fault

Check all connections and wiring. Contact

 

 

 

 

 

Invacare Technical Service.

 

 

 

 

 

 

11

E11

Incompati-

00

Incompatible/incorrect

Wrong type of remote connected. Ensure

 

 

ble/incorrect

 

Remote

the branding of the joystick matches that

 

 

Remote

 

 

of controller unit.

 

 

 

 

 

 

Motor/Gearbox/Brake

 

 

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

 

 

 

 

Bad bearings.

If bearings are bad, replace motor. Refer to

 

 

Removing/Installing the Motor on page 112.

Motor makes a

 

 

Raised commutator plate inside of

Ohm out motor and replace motor if high reading

clicking noise.

motor.

is present. Normal reading is .5-5 Ohms. Refer to

 

 

 

Motor Testing on page 31 or Removing/Installing

 

 

the Motor on page 112.

 

 

 

Grinding noise or

Bad bearings. Bad Gears.

Replace motor/gearbox. Refer to

motor is locking

 

Removing/Installing the Motor on page 112.

up.

 

 

 

 

 

Motors stall and

Current Rollback.

Stop driving and let electronics cool.

starts up again.

 

 

 

 

 

Part No. 1125075

25

Pronto® Series

Page 25
Image 25
Invacare M51, M50, M61 service manual Motor/Gearbox/Brake, Symptom Probable Cause Solutions